Appendix B: Calibration
2606B System SourceMeter® Instrument Reference Manual
B-22
2606B-901-01 Rev. B / May 2018
Typical values for the 100 mA range:
smua.source.calibratei(100e-3, 1e-30, 1e-5, 90e-3, 88e-3)
12. If this step is not on the CALA sense mode, send the measure calibration command using the
multimeter and 2606B readings, and range setting for the parameters:
smua.measure.calibratei(
range, Z_rdg, DMM_Z_rdg, FS_rdg, DMM_FS_rdg
)
Where:
range
= The present calibration range
Z_rdg
=
+zero 2606B measurement
DMM_Z_rdg
= The +zero DMM measurement
FS_rdg
=
+FS 2606B measurement
DMM_FS_rdg
= The +FS DMM measurement
Typical 100 mA range values:
smua.measure.calibratei(100e-3, 1e-6, 1e-5, 0.089, 0.088)
13. Select negative polarity, then set the source output to the negative zero value, for example:
smua.cal.polarity = smua.CAL_NEGATIVE
smua.source.leveli = -1e-30
14. Turn on the output:
smua.source.output = smua.OUTPUT_ON
15. Allow the readings to settle, then get both the multimeter and 2606B current readings at the
negative zero value (the 2606B measurement is not necessary if this calibration step is being
done on the CALA sense). The two measurements should be made as close as possible in time.
Use this command for the 2606B:
Z_rdg = smua.measure.i()
16. Turn off the output:
smua.source.output = smua.OUTPUT_OFF
17. Set the source output to the negative full-scale value, for example:
smua.source.leveli = -90e-3
18. Turn on the output:
smua.source.output = smua.OUTPUT_ON
19. Allow the readings to settle, then get both the multimeter and 2606B current readings at the
negative full-scale output value (the 2606B measurement is not necessary if this calibration step
is being done on the CALA sense mode). The two measurements should be made as close as
possible in time. Use this command for the 2606B:
FS_rdg = smua.measure.i()
20. Turn off the output:
smua.source.output = smua.OUTPUT_OFF
21. Send the source calibration command using the -range, -zero and -FS multimeter readings, and -
zero and -FS source values for the parameters:
smua.source.calibratei(
-range, src_Z, DMM_Z_rdg, src_FS, DMM_FS_rdg
)
Where:
-range
= The negative of the present calibration range
src_Z
= The zero 2606B source output value
DMM_Z_rdg
= The zero DMM measurement
src_FS
= The FS 2606B source output value
DMM_FS_rdg
= The FS DMM measurement