Section 7: TSP command reference
2606B System SourceMeter® Instrument Reference Manual
7-130
2606B-901-01 Rev. B / May 2018
Also see
(on page 7-120)
(on page 7-130)
(on page 7-133)
(on page 7-134)
(on page 7-136)
(on page 7-138)
Understanding hardware value and pseudo line state
(on page 3-48)
lan.trigger[N].clear()
This function clears the event detector for a LAN trigger.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
lan.trigger[
N
].clear()
N
The LAN event number (1 to 8) to clear
Details
The trigger event detector enters the detected state when an event is detected. This function clears a
trigger event detector and discards the previous history of the trigger packet.
This function clears all overruns associated with this LAN trigger.
Example
lan.trigger[5].clear()
Clears the event detector with LAN packet 5.
Also see
(on page 7-129)
(on page 7-134)
(on page 7-136)
(on page 7-138)
lan.trigger[N].connect()
This function prepares the event generator for outgoing trigger events.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Function
Yes
Usage
lan.trigger[
N
].connect()
N
The LAN event number (1 to 8)
Details
This command prepares the event generator to send event messages. For TCP connections, this
opens the TCP connection.