Section 7: TSP command reference
2606B System SourceMeter® Instrument Reference Manual
7-258
2606B-901-01 Rev. B / May 2018
As an example, to set bit B1 of the measurement event instrument summary enable register, set
status.measurement.instrument.enable = status.measurement.instrument.SMUA
.
In addition to the above constants,
measurementRegister
can be set to the decimal equivalent of
the bit to set. To set more than one bit of the register, set
measurementRegister
to the sum of
their decimal weights. For example, to set bits B1 and B2, set
measurementRegister
to 6 (which is
the sum of 2 + 4).
Example
status.measurement.instrument.enable =
status.measurement.instrument.SMUA
Sets the SMU A bit of the
measurement event instrument
summary enable register using a
constant.
Also see
(on page E-21)
status.measurement.instrument.smuX.*
This attribute contains the registers of the measurement event SMU X summary register set.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute
- -
- -
- -
- -
.condition (R)
Yes
Not applicable
Not saved
Not applicable
.enable (RW)
Yes
Status reset
Not saved
0
.event (R)
Yes
Status reset
Not saved
0
.ntr (RW)
Yes
Status reset
Not saved
0
.ptr (RW)
Yes
Status reset
Not saved
387 (All bits set)
Usage
measurementRegister
= status.measurement.instrument.smu
X
.condition
measurementRegister
= status.measurement.instrument.smu
X
.enable
measurementRegister
= status.measurement.instrument.smu
X
.event
measurementRegister
= status.measurement.instrument.smu
X
.ntr
measurementRegister
= status.measurement.instrument.smu
X
.ptr
status.measurement.instrument.smu
X
.enable =
measurementRegister
status.measurement.instrument.smu
X
.ntr =
measurementRegister
status.measurement.instrument.smu
X
.ptr =
measurementRegister
measurementRegister
The status of the instrument measurement status SMU X summary register; a
zero (0) indicates no bits set (also send 0 to clear all bits); other values indicate
various bit settings
X
Source-measure unit (SMU) channel (for example
status.measurement.instrument.smua.enable
applies to SMU
channel A)
Details
These attributes are used to read or write to the measurement event SMU X summary registers.
Reading a status register returns a value. The binary equivalent of the returned value indicates which
register bits are set. The least significant bit of the binary number is bit B0, and the most significant bit
is bit B15. For example, assume the value
257
is returned for the enable register. The binary
equivalent is
0000 0001 0000 0001
. This value indicates that bit B0 (VLMT) and bit B8 (BAV) are
set.