Section 7: TSP command reference
2606B System SourceMeter® Instrument Reference Manual
7-232
2606B-901-01 Rev. B / May 2018
smuX.trigger.endpulse.stimulus
This attribute defines which event will cause the end pulse event detector to enter the detected state.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
SMU reset
Instrument reset
Recall setup
Not saved
0
Usage
eventID
= smu
X
.trigger.endpulse.stimulus
smu
X
.trigger.endpulse.stimulus =
eventID
eventID
Set to the event that triggers the end pulse action of the source
X
Source-measure unit (SMU) channel (for example,
smua.trigger.endpulse.stimulus
applies to SMU channel A)
Details
Set this attribute to the event ID of any trigger event generator to wait for that event. To bypass
waiting for an event, set this attribute's value to
0
. Set
eventID
to one of the existing trigger event
IDs, which are shown in the following table.
Trigger event IDs*
Event ID
Event description
smu
X
.trigger.SWEEPING_EVENT_ID
Occurs when the source-measure unit (SMU) transitions
from the idle state to the arm layer of the trigger model
smu
X
.trigger.ARMED_EVENT_ID
Occurs when the SMU moves from the arm layer to the
trigger layer of the trigger model
smu
X
.trigger.SOURCE_COMPLETE_EVENT_ID
Occurs when the SMU completes a source action
smu
X
.trigger.MEASURE_COMPLETE_EVENT_ID
Occurs when the SMU completes a measurement action
smu
X
.trigger.PULSE_COMPLETE_EVENT_ID
Occurs when the SMU completes a pulse
smu
X
.trigger.SWEEP_COMPLETE_EVENT_ID
Occurs when the SMU completes a sweep
smu
X
.trigger.IDLE_EVENT_ID
Occurs when the SMU returns to the idle state
digio.trigger[
N
].EVENT_ID
Occurs when an edge is detected on a digital I/O line
tsplink.trigger[
N
].EVENT_ID
Occurs when an edge is detected on a TSP-Link line
lan.trigger[
N
].EVENT_ID
Occurs when the appropriate LXI trigger packet is received
on LAN trigger object
N
display.trigger.EVENT_ID
Occurs when the TRIG key on the virtual front panel is
pressed
trigger.EVENT_ID
Occurs when a
*TRG
command is received on the remote
interface
USB only: Occurs when a USBTMC TRIGGER message is
received
VXI-11 only: Occurs with the VXI-11 command
device_trigger
; reference the VXI-11 standard for
additional details on the device trigger operation
trigger.blender[
N
].EVENT_ID
Occurs after a collection of events is detected