Section 7: TSP command reference
2606B System SourceMeter® Instrument Reference Manual
7-240
2606B-901-01 Rev. B / May 2018
Details
Set the stimulus of any trigger object to the value of this constant to have the trigger object respond to
pulse complete events from this SMU.
Also see
(on page 3-30)
smuX.trigger.source.action
This attribute enables or disables sweeping the source (on or off).
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute (RW)
Yes
SMU reset
Instrument reset
Recall setup
Not saved
0 (smuX.DISABLE)
Usage
action
= smu
X
.trigger.source.action
smu
X
.trigger.source.action =
action
action
Sweep source action. Set to one of the following values:
0
or
smu
X
.DISABLE
: Do not sweep the source
1
or
smu
X
.ENABLE
: Sweep the source
X
Source-measure unit (SMU) channel (for example,
smua.trigger.source.action
applies to SMU channel A)
Details
This attribute is used to enable or disable source level changes during a sweep. In addition to
enabling the action before initiating the sweep, make sure to configure it using
smu
X
.trigger.source.linear
Y
()
,
smu
X
.trigger.source.list
Y
()
, or
smu
X
.trigger.source.log
Y
()
.
Example
smua.trigger.source.listv({3, 1, 4, 5, 2})
smua.trigger.source.action = smua.ENABLE
Configure list sweep for SMU channel A (sweep through 3 V, 1 V, 4 V, 5 V, and 2 V).
Enable the source action.
Also see
(on page 7-243)
(on page 7-244)
(on page 7-245)
(on page 3-30)