Section 7: TSP command reference
2606B System SourceMeter® Instrument Reference Manual
7-268
2606B-901-01 Rev. B / May 2018
As an example, to set bit B12 of the operation status enable register, set
status.operation.enable
= status.operation.USER
.
In addition to the above constants,
operationRegister
can be set to the numeric equivalent of the
bit to set. To set more than one bit of the register, set
operationRegister
to the sum of their
decimal weights. For example, to set bits B12 and B14, set
operationRegister
to 20,480 (which
is the sum of 4,096 + 16,384).
Bit
B7
B6
B5
B4
B3
B2
B1
B0
Binary value
0/1
0/1
0/1
0/1
0/1
0/1
0/1
0/1
Decimal
128
64
32
16
8
4
2
1
Weights
(2
7
)
(2
6
)
(2
5
)
(2
4
)
(2
3
)
(2
2
)
(2
1
)
(2
0
)
Bit
B15
B14
B13
B12
B11
B10
B9
B8
Binary value
0/1
0/1
0/1
0/1
0/1
0/1
0/1
0/1
Decimal
32,768
16,384
8,192
4,096
2,048
1,024
512
256
Weights
(2
15
)
(2
14
)
(2
13
)
(2
12
)
(2
11
)
(2
10
)
(2
9
)
(2
8
)
Example 1
operationRegister = status.operation.USER + status.operation.PROG
status.operation.enable = operationRegister
Sets the USER and PROG bits of the operation status enable register using constants.
Example 2
-- decimal 20480 = binary 0101 0000 0000 0000
operationRegister = 20480
status.operation.enable = operationRegister
Sets the USER and PROG bits of the operation status enable register using a decimal value.
Also see
(on page E-7)
status.operation.calibrating.*
This attribute contains the operation status calibration summary register set.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Attribute
- -
- -
- -
- -
.condition (R)
Yes
Not applicable
Not saved
Not applicable
.enable (RW)
Yes
Status reset
Not saved
0
.event (R)
Yes
Status reset
Not saved
0
.ntr (RW)
Yes
Status reset
Not saved
0
.ptr (RW)
Yes
Status reset
Not saved
6 (All bits set)