Section 8: Troubleshooting guide
2606B System SourceMeter® Instrument Reference Manual
8-6
2606B-901-01 Rev. B / May 2018
Error summary
Error number
Error level
Error message
5044
RECOVERABLE
Cannot perform requested operation while current measure
autorange is enabled
5045
RECOVERABLE
Cannot perform requested operation while filter is enabled
5046
SERIOUS
SMU too hot
5047
RECOVERABLE
Minimum timestamp resolution is 1us
5048
RECOVERABLE
Contact check not valid with HIGH-Z OUTPUT off
5049
RECOVERABLE
Contact check not valid while an active current source
5050
RECOVERABLE
I limit too low for contact check
5051
FATAL
Model number/SMU hardware mismatch. Disconnect DUT and cycle
power
5053
RECOVERABLE
Unstable output detected - Measurements may not be valid
5055
RECOVERABLE
Cannot change adjustment date - change cal constants before
operation
5059
RECOVERABLE
trigger.source.action enabled without configuration
5060
RECOVERABLE
trigger.measure.action enabled without configuration
5061
RECOVERABLE
Operation not permitted while OUTPUT is off
5062
SERIOUS
SMU overload. Automatic OUTPUT off.
5063
RECOVERABLE
Cannot perform requested operation while measure autozero is on
5064
RECOVERABLE
Cannot use reading buffer that collects source values
5065
RECOVERABLE
I range too low for contact check
5066
RECOVERABLE
source.offlimiti too low for contact check
5069
SERIOUS
Autorange locked for HighC mode
LAN troubleshooting suggestions
If you are unable to connect to the instrument's web interface, check the following items:
•
Verify that the network cable is in the LAN port on the rear panel of the instrument, not one of the
TSP-Link
®
(on page 2-3)).
•
Verify that the network cable is in the correct port on the computer. The LAN port of a laptop may
be disabled when the laptop is in a docking station.
•
Verify that the correct ethernet card's configuration information was used during the setup
procedure.
•
Verify that the computer's network card is enabled.
•
Verify that the instrument's IP address is compatible with the IP address on the computer.
•
Verify that the instrument's subnet mask address is the same as the computer's subnet mask
address.
•
Turn the instrument's power off, and then on. Wait at least 60 seconds for the network
configuration to be completed. Verify that an IP address has been assigned to the instrument:
1. Press the
MENU
key to display the MAIN MENU.
2. Use the navigation wheel to select
LAN
. The LAN CONFIG menu is displayed.
3. Select
STATUS
.
4. Select
IP-ADDRESS
.
•
Restart your computer.
•
For more detail on LAN settings, see
(on page C-2).
If the above actions do not correct the problem, contact your system administrator.