Example
smua.trigger.measure.v(vbuffername)
smua.trigger.measure.action = smua.ENABLE
Stores voltage readings during the
sweep for SMU channel A in buffer
vbuffername
.
Also see
(on page 7-209)
(on page 7-235)
(on page 3-30)
(on page 7-383)
smuX.trigger.MEASURE_COMPLETE_EVENT_ID
This constant contains the measurement complete event number.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Constant
Yes
Usage
eventID
= smu
X
.trigger.MEASURE_COMPLETE_EVENT_ID
eventID
The measurement complete event number
X
Source-measure unit (SMU) channel (for example,
smua.trigger.MEASURE_COMPLETE_EVENT_ID
applies to SMU channel A)
Details
Set the stimulus of any trigger object to the value of this constant to have the trigger object respond to
measure complete events from this SMU.
Also see
(on page 3-30)
smuX.trigger.PULSE_COMPLETE_EVENT_ID
This constant contains the pulse complete event number.
Type
TSP-Link accessible
Affected by
Where saved
Default value
Constant
Yes
Usage
eventID
= smu
X
.trigger.PULSE_COMPLETE_EVENT_ID
eventID
The pulse complete event number
X
Source-measure unit (SMU) channel (for example,
smua.trigger.PULSE_COMPLETE_EVENT_ID
applies to SMU channel A)