![Tektronix 2440 User Manual Download Page 425](http://html1.mh-extra.com/html/tektronix/2440/2440_user-manual_1078860425.webp)
TEST WAVEFORMS
Test waveforms for schematic diagrams (if applicable) precede the schematic. They are intended to aid in trouble
shooting the instrument. Special conditions required of the test oscilloscope are given above the waveform; special condi
tions for the 2440 under test are given beneath the waveform. Unless otherwise stated, the test conditions for the first
waveform listed pertain to all the waveforms for a given schematic diagram. Normal control settings for the test oscillo
scope are given in the readouts shown in each waveform illustration.
RECOMMENDED TEST EQUIPMENT
Item
Specification
Example
Test oscilloscope with 10X probe
and 1X probe (1X probe is optional
accessory.)
Digital Storage; frequency response:
dc to 40 MHz single event bandwidth
to 150 MHz equivalent-time sampling.
TEKTRONIX 2440 Digital Oscilloscope
with two 10X probes.
Calibration Generator
Standard-amplitude
signal
levels:
5 mV to 50 V. Accuracy ± 3 % .
TEKTRONIX PG 506 Calibration
Generator.®
Digital Voltmeter (DMM)
Range 0 to 140 V. Dc voltage accu
racy: ± 0 .1 5 % . 4 1/2 digit display.
TEKTRONIX DM 501A Digital
Multimeter.®
High Voltage Probe for DMM
Maximum voltage 20 kV.
Fluke Model 80K-40 High Voltage
Probe.
Precision Coaxial Cable
Impedance: 50 ohm. Length: 36 in.
Connectors: BNC.
Tektronix Part Number 012-0482-00.
Dual-Input Coupler
Connectors: BNC female-to-dual-BNC
male.
Tektronix Part Number 103-0090-00.
Sync and Linearity Test Generator
Conforms to TV System require
ments.
TEKTRONIX R147A NTSC Test Sig
nal Generator.
TEKTRONIX R 148 Insertion Test Sig
nal Generator.
Coaxial Cable (2 required)
Impedance: 75 ohm. Length: 42 in.
Connectors: BNC.
Tektronix Part Number 012-0074-00.
aTM 500-Series Power-Module Mainframe required.
DC VOLTAGE MEASUREMENTS
Dc voltages indicated on the schematic diagrams are typical of a normally operating instrument. Voltages are with
respect to chassis ground except in the isolated portion of the Low Voltage Power Supply, where they are with respect to
the REF NODE indicated by the darker line in the Control Power Supply circuitry.
Summary of Contents for 2440
Page 4: ......
Page 12: ......
Page 38: ......
Page 91: ...Theory of Operation 2440 Service 3 49 Figure 3 5 Simplified CCD architecture ...
Page 120: ...Theory of Operation 2440 Service Figure 3 10 DC Restorer 3 78 ...
Page 130: ...Theory of Operation 2440 Service 3 88 ...
Page 136: ...Theory of Operation 2440 Service Figure 3 15 PWM Regulator and Inverter 3 94 ...
Page 138: ...Theory of Operation 2440 Service Figure 3 16 PWM switching waveforms 3 96 ...
Page 208: ......
Page 322: ......
Page 419: ......
Page 423: ...2440 Service Figure 9 4b 2440 Block diagram part 2 6330 27 BLOCK DIAGRAM PART 2 ...
Page 427: ......
Page 428: ...2440 Service WAVEFORMS FOR DIAGRAM 1 6 2 8 5 3 2 ...
Page 429: ......
Page 432: ...2440 Service WAVEFORMS FOR DIAGRAM 2 BUS ISOLATED 6 6 0 3 4 9 ...
Page 433: ...WAVEFORMS FOR DIAGRAM 2 ...
Page 439: ......
Page 445: ......
Page 450: ...2440 Service WAVEFORMS FOR DIAGRAM 8 6603 27 ...
Page 455: ...W A V EFO R M S FOR D IAG RAM 11 WAVEFORMS FOR DIAGRAM 11 6603 29 ...
Page 458: ...WAVEFORMS FOR DIAGRAM 12 ...
Page 463: ...WAVEFORMS FOR DIAGRAM 14 ...
Page 471: ... C N i u i t e x r 4 j u J l i 1 i 6 28 5 44 WAVEFORMS FOR DIAGRAM 18 ...
Page 476: ...WAVEFORMS FOR DIAGRAM 20 ...
Page 490: ...YES ...
Page 492: ......
Page 494: ......
Page 499: ...YES ...
Page 501: ......
Page 502: ...2440 Service FIG 1 C A B IN E T ...
Page 503: ......
Page 510: ...2440 Service SISSVHO Z Old ...
Page 511: ......
Page 512: ...A12 2440 Service F IG 3 C IR C U IT BOARDS ...
Page 513: ......
Page 517: ...10 16 R E V J U L 1993 Replaceable Mechanical Parts ...
Page 518: ...2440 Service ...
Page 519: ...FIG 4 L V H V PWR SUPPLY ...
Page 520: ...FIG 5 ACCESSORIES 2440 Service ...
Page 521: ......
Page 523: ......
Page 529: ......
Page 530: ......
Page 531: ......