Perform ance Check and Functional Verification Procedure
2440 Service
f.
Push the front-panel button labeled AUTO to do an Auto
Setup on the input waveform for CH 1.
g.
VERIFY - That the scope displays the Auto Setup menu
and the message “AUTOSETUP WORKING: PLEASE
W AIT’ as it acquires information about the CH 1
waveform.
h.
V E R IF Y -T h a t the Auto Setup mode is VIEW (from the
recalled front-panel setup).
i.
VERIFY (after the message is re m o v e d )-T h a t the
ground-reference indicator for the waveform (“ + *) is
approximately aligned to the horizontal graticule line at
center screen, and that the entire waveform is con
tained vertically within the graticule area. (The trigger
point indicator, a small “T ” riding on the waveform,
should be approximately at horizontal center screen.)
j.
Vertically reposition the ground-reference indicator to
2 divisions below the center horizontal graticule line,
then return the CH 1 VOLTS/DIV to 20 mV (same set up
as part e).
k.
Select VERTICAL M O DE and set CH 2 on. Then,
disconnect the test setup at CH 1 and reconnect the
Pulse Generator to both CH 1 OR X and CH 2 OR Y
input connectors through the 50 O Coaxial Cable,
followed by the 10X Attenuator, followed by a
Dual-Input Coupler.
l
.
Now, position the ground-reference indicator for CH 2
to 2 divisions above the center horizontal graticule line;
then set CH 2 VOLTS/DIV to 20 mV.
m. Push AUTO. V E R IF Y -T h a t the scope scales both the
C H 1 a n d C H 2 waveforms so that the CH 1 waveform is
contained within the top half of the screen and the CH 2
waveform is contained within the bottom half. (Slightly
adjust the vertical position knobs to see which is the
CH 1 waveform and which is the CH 2 waveform.)
n.
Set PERIOD on in the Auto Setup menu. V ER IFY—That
the menu entry RES HI | LO appears with the setting LO
selected.
o.
Disconnect the test setup at CH 1 OR X and CH 2 OR Y
input connectors, remove the Dual-Input Coupler from
the setup, and reconnect the setup to the CH 1 OR X
input connector. (Same connection as in part b.)
p.
Select VERTICAL M O D E and 6et CH 2 off.
q.
Push the AUTO button. V E R IF Y -T h a t between 1 and 2
cycles of the waveform are displayed on screen, with
the amplitude contained within about ± 2 .2 5 divisions
of center. (The amplitude is about 2.5 divisions for this
particular input waveform.) The trigger-point indicator
is near the beginning of the 20-division waveform
record, and the Trigger SLOPE is positive (“ + ").
r.
Set RES HI | LO to HI in the Auto Setup menu.
s.
Push AUTO. V E R IF Y -T h a t the waveform is displayed
with higher vertical resolution (more amplitude than in
part q, about 5 divisions for this particular input
waveform) and that the amplitude is contained
vertically on screen. Also verify that about 1-2 cycles is
included in the ENTIRE 20-division waveform record.
Use the HORIZONTAL POSITION control to view the
entire waveform. The trigger-point indicator is near the
beginning of the record; the Trigger SLOPE is
positive (“ + ’ ).
t.
Set RES HI | LO to LO and PULSE on in the Auto Setup
menu.
u.
Push AUTO. V E R IF Y -T h a t the positive V
4
-cycle of the
waveform (that is, the pulse) is displayed on screen and
that the amplitude is contained vertically within approx
imately ± 2 .2 5 divisions of center screen. (The am pli
tude is about 2.5 divisions for this particular input
waveform.) The trigger-point indicator is near the
beginning of the 20-division waveform record, and the
Trigger SLOPE is positive (“ + ").
v.
Set the A S E C /D IV to 20 ns, then set the Pulse
Generator to produce a C O M PLEM EN T pulse; that is,
one with a negative V
4
-cycle pulse duration.
w.
Push AUTO. V E R IF Y -T h a t the negative V
4
-cycle of the
waveform (in this case the pulse is negative-going) is
displayed on screen and that the amplitude is con
tained vertically within approximately ± 2 .2 5 divisions
of center screen. (The amplitude is about 2.5 divisions
for this particular input waveform.) T he trigger point
indicator is near the beginning of the record; the Trigger
SLOPE is negative (“-* ).
x.
Set the RES HI | LO to HI.
y.
Push AUTO. VERIFY - That the negative V
4
-cycle of the
waveform is displayed over about 10 of the 20 divisions
in the waveform record, with the amplitude contained
vertically on screen. (The amplitude is about 5 divisions
for this particular input.) The trigger point indicator is
near the beginning of the record; the Trigger SLOPE is
negative ( * - ”).
z .
S et the
RES
H I|L O
to
LO
and
the
EDG E
- / " | " \ _ t o j r ,
aa. Push AUTO. V E R IF Y -T h a t the rising (positive-going)
edge of the waveform is displayed on screen with the
trigger-point indicator at center screen. (Waveform
amplitude is about 2.5 divisions for this particular
input waveform).
ab. Set EDGE
- T \
“ V - to
ac. Push AUTO. V E R IF Y -T h a t the falling (negative-going)
edge of the waveform is displayed on screen with the
trigger-point at center screen. (The amplitude is about
2.5 divisions for this particular Input waveform).
4-30
Summary of Contents for 2440
Page 4: ......
Page 12: ......
Page 38: ......
Page 91: ...Theory of Operation 2440 Service 3 49 Figure 3 5 Simplified CCD architecture ...
Page 120: ...Theory of Operation 2440 Service Figure 3 10 DC Restorer 3 78 ...
Page 130: ...Theory of Operation 2440 Service 3 88 ...
Page 136: ...Theory of Operation 2440 Service Figure 3 15 PWM Regulator and Inverter 3 94 ...
Page 138: ...Theory of Operation 2440 Service Figure 3 16 PWM switching waveforms 3 96 ...
Page 208: ......
Page 322: ......
Page 419: ......
Page 423: ...2440 Service Figure 9 4b 2440 Block diagram part 2 6330 27 BLOCK DIAGRAM PART 2 ...
Page 427: ......
Page 428: ...2440 Service WAVEFORMS FOR DIAGRAM 1 6 2 8 5 3 2 ...
Page 429: ......
Page 432: ...2440 Service WAVEFORMS FOR DIAGRAM 2 BUS ISOLATED 6 6 0 3 4 9 ...
Page 433: ...WAVEFORMS FOR DIAGRAM 2 ...
Page 439: ......
Page 445: ......
Page 450: ...2440 Service WAVEFORMS FOR DIAGRAM 8 6603 27 ...
Page 455: ...W A V EFO R M S FOR D IAG RAM 11 WAVEFORMS FOR DIAGRAM 11 6603 29 ...
Page 458: ...WAVEFORMS FOR DIAGRAM 12 ...
Page 463: ...WAVEFORMS FOR DIAGRAM 14 ...
Page 471: ... C N i u i t e x r 4 j u J l i 1 i 6 28 5 44 WAVEFORMS FOR DIAGRAM 18 ...
Page 476: ...WAVEFORMS FOR DIAGRAM 20 ...
Page 490: ...YES ...
Page 492: ......
Page 494: ......
Page 499: ...YES ...
Page 501: ......
Page 502: ...2440 Service FIG 1 C A B IN E T ...
Page 503: ......
Page 510: ...2440 Service SISSVHO Z Old ...
Page 511: ......
Page 512: ...A12 2440 Service F IG 3 C IR C U IT BOARDS ...
Page 513: ......
Page 517: ...10 16 R E V J U L 1993 Replaceable Mechanical Parts ...
Page 518: ...2440 Service ...
Page 519: ...FIG 4 L V H V PWR SUPPLY ...
Page 520: ...FIG 5 ACCESSORIES 2440 Service ...
Page 521: ......
Page 523: ......
Page 529: ......
Page 530: ......
Page 531: ......