![Tektronix 2440 User Manual Download Page 35](http://html1.mh-extra.com/html/tektronix/2440/2440_user-manual_1078860035.webp)
Specification— 2440 Service
Table 1-4
Option 05 (TV Trigger) Electrical Characteristics
Characteristics
Performance Requirements
VERTICAL— CHANNEL 1 AND CHANNEL 2
Frequency Response
Full Bandwidth
50 kHz to 5 MHz
Within ± 1 % .
Greater than 5 MHz to 10 MHz
Within + 1 % ,
2%.
Greater than 10 MHz to 30 MHz
Within + 2 % , - 3 % .
For VOLTS/DIV switch settings between 5 mV and 0.2 V per divi
sion with VARIABLE VOLTS/DIV set to CAL. Five-division,
50 kHz reference signals from a 50 Si system. With external 50 St
termination on a 1 MS2 input.
20 MHz Bandwidth Limit
50 kHz to 5 MHz
Within + 1 % , - 4 % .
Square Wave Flatness
Field Rate
5 mV/div to 20 mV/div
± 1 % , 1% p-p at 60 Hz with input signal of 0.1 V.
50 mV/div
± 1 % , 1% p-p at 60 Hz with input signal of 1.0 V.
With fast-rise step (rise time 1 ns or less), 1 MS2 dc input coupling,
an external 50 St termination, and VARIABLE VOLTS/DIV set to
CAL. Exclude the first 20 ns following the step transition and
exclude the first 30 ns when 20 MHz BW LIMIT is set.
Line Rate
5 mV/div to 20 mV/div
± 1 % , 1% p-p at 15 kHz with input signal of 0.1 V.
50 mV/div
± 1 %, 1 % p-p at 15 kHz with input signal of 1.0 V.
TV (Back-Porch) Clamp (CH 2 Only)
60 Hz Attenuation
18 dB or greater.
For VOLTS/DIV switch settings between 5 mV and 0.2 V with
VARIABLE VO LTS/DIV set to CAL. Six-division reference signal.
Back-Porch Reference
Within ± 1 .0 division of ground reference.
1-23
Summary of Contents for 2440
Page 4: ......
Page 12: ......
Page 38: ......
Page 91: ...Theory of Operation 2440 Service 3 49 Figure 3 5 Simplified CCD architecture ...
Page 120: ...Theory of Operation 2440 Service Figure 3 10 DC Restorer 3 78 ...
Page 130: ...Theory of Operation 2440 Service 3 88 ...
Page 136: ...Theory of Operation 2440 Service Figure 3 15 PWM Regulator and Inverter 3 94 ...
Page 138: ...Theory of Operation 2440 Service Figure 3 16 PWM switching waveforms 3 96 ...
Page 208: ......
Page 322: ......
Page 419: ......
Page 423: ...2440 Service Figure 9 4b 2440 Block diagram part 2 6330 27 BLOCK DIAGRAM PART 2 ...
Page 427: ......
Page 428: ...2440 Service WAVEFORMS FOR DIAGRAM 1 6 2 8 5 3 2 ...
Page 429: ......
Page 432: ...2440 Service WAVEFORMS FOR DIAGRAM 2 BUS ISOLATED 6 6 0 3 4 9 ...
Page 433: ...WAVEFORMS FOR DIAGRAM 2 ...
Page 439: ......
Page 445: ......
Page 450: ...2440 Service WAVEFORMS FOR DIAGRAM 8 6603 27 ...
Page 455: ...W A V EFO R M S FOR D IAG RAM 11 WAVEFORMS FOR DIAGRAM 11 6603 29 ...
Page 458: ...WAVEFORMS FOR DIAGRAM 12 ...
Page 463: ...WAVEFORMS FOR DIAGRAM 14 ...
Page 471: ... C N i u i t e x r 4 j u J l i 1 i 6 28 5 44 WAVEFORMS FOR DIAGRAM 18 ...
Page 476: ...WAVEFORMS FOR DIAGRAM 20 ...
Page 490: ...YES ...
Page 492: ......
Page 494: ......
Page 499: ...YES ...
Page 501: ......
Page 502: ...2440 Service FIG 1 C A B IN E T ...
Page 503: ......
Page 510: ...2440 Service SISSVHO Z Old ...
Page 511: ......
Page 512: ...A12 2440 Service F IG 3 C IR C U IT BOARDS ...
Page 513: ......
Page 517: ...10 16 R E V J U L 1993 Replaceable Mechanical Parts ...
Page 518: ...2440 Service ...
Page 519: ...FIG 4 L V H V PWR SUPPLY ...
Page 520: ...FIG 5 ACCESSORIES 2440 Service ...
Page 521: ......
Page 523: ......
Page 529: ......
Page 530: ......
Page 531: ......