![Tektronix 2440 User Manual Download Page 291](http://html1.mh-extra.com/html/tektronix/2440/2440_user-manual_1078860291.webp)
Maintenance— 2440 Service
Table 6-6 (cont)
5300
VERSION-CHK
Waveform
n P
ROM A12U480 and A12U490 (schematic diagram 2):
Testing Method:
The version number in the header is preset to “ ?" and is filled in by this test. If the test fails, the “?"
will remain in the header for further indication of an error. A Waveform mP reset causes the Waveform
mP to read the version number bytes of the Waveform MP code. If the version number is incorrect, the
Waveform
» P
code is incompatible with the System
fiP
code and may not execute properly.
Troubleshooting Procedure:
If test 5300 fails, replace Waveform
n P
ROMs U480 and/or U490 with the correct ones for the version
of System
n
P code being used.
6000
CK SUM -NVRAM
Nonvolatile RAM Checksum A12U664 (schematic diagram 1):
Testing Method:
Some of the CRCCs (check sums) are computed at power-down and will be valid only at power-up.
Therefore, executing tests 5000 through 5003 will only display the flags that resulted from power-up
diagnostics.
N O T E
F A IL a n d P A S S fla g s in th e E x te n d e d D ia g n o s tic s m e n u s h o w th e re s u lts o f th e la s t te s t
ra n . I f a d e fe c tiv e d e v ic e th a t h a s p re v io u s ly c a u s e d a F A IL fla g to b e s e t is re p la c e d , th e
te s t m u s t b e ru n a g a in to o b ta in a P A S S in d ic a tio n in th e m enu.
When the instrument is SELF CALIBRATED, a CRCC is calculated and stored for the Calibration Con
stants in NV RAM.
When power-down is executed, the values of the front-panel variables have a CRCC calculated and
stored.
When a waveform is saved, the CRCC is calculated for the waveform and headers and saved.
On power-up, all of these are recalculated and compared to the stored CRCC word. If they do not
agree, that test fails.
6100
CAL
CONSTANTS
Calibration Constants:
Troubleshooting Procedure:
If FAIL, the calibration constants have been lost and a COLD START is executed. The instrument
must be recalibrated to return to calibrated operation after a COLD START.
A failure of 6100 is serious to the normal operation of the scope, and the cause of the failure
should be found and corrected to prevent reoccurrence.
1.
Check BT800 and the components that connect and disconnect the battery from the NV RAM at
power-off and power-on respectively.
6 -8 3
Summary of Contents for 2440
Page 4: ......
Page 12: ......
Page 38: ......
Page 91: ...Theory of Operation 2440 Service 3 49 Figure 3 5 Simplified CCD architecture ...
Page 120: ...Theory of Operation 2440 Service Figure 3 10 DC Restorer 3 78 ...
Page 130: ...Theory of Operation 2440 Service 3 88 ...
Page 136: ...Theory of Operation 2440 Service Figure 3 15 PWM Regulator and Inverter 3 94 ...
Page 138: ...Theory of Operation 2440 Service Figure 3 16 PWM switching waveforms 3 96 ...
Page 208: ......
Page 322: ......
Page 419: ......
Page 423: ...2440 Service Figure 9 4b 2440 Block diagram part 2 6330 27 BLOCK DIAGRAM PART 2 ...
Page 427: ......
Page 428: ...2440 Service WAVEFORMS FOR DIAGRAM 1 6 2 8 5 3 2 ...
Page 429: ......
Page 432: ...2440 Service WAVEFORMS FOR DIAGRAM 2 BUS ISOLATED 6 6 0 3 4 9 ...
Page 433: ...WAVEFORMS FOR DIAGRAM 2 ...
Page 439: ......
Page 445: ......
Page 450: ...2440 Service WAVEFORMS FOR DIAGRAM 8 6603 27 ...
Page 455: ...W A V EFO R M S FOR D IAG RAM 11 WAVEFORMS FOR DIAGRAM 11 6603 29 ...
Page 458: ...WAVEFORMS FOR DIAGRAM 12 ...
Page 463: ...WAVEFORMS FOR DIAGRAM 14 ...
Page 471: ... C N i u i t e x r 4 j u J l i 1 i 6 28 5 44 WAVEFORMS FOR DIAGRAM 18 ...
Page 476: ...WAVEFORMS FOR DIAGRAM 20 ...
Page 490: ...YES ...
Page 492: ......
Page 494: ......
Page 499: ...YES ...
Page 501: ......
Page 502: ...2440 Service FIG 1 C A B IN E T ...
Page 503: ......
Page 510: ...2440 Service SISSVHO Z Old ...
Page 511: ......
Page 512: ...A12 2440 Service F IG 3 C IR C U IT BOARDS ...
Page 513: ......
Page 517: ...10 16 R E V J U L 1993 Replaceable Mechanical Parts ...
Page 518: ...2440 Service ...
Page 519: ...FIG 4 L V H V PWR SUPPLY ...
Page 520: ...FIG 5 ACCESSORIES 2440 Service ...
Page 521: ......
Page 523: ......
Page 529: ......
Page 530: ......
Page 531: ......