![Tektronix 2440 User Manual Download Page 167](http://html1.mh-extra.com/html/tektronix/2440/2440_user-manual_1078860167.webp)
l
.
Set the Tim e Mark Generator’s marker period to match
the acquisition rate set in the last part.
m. V ER IFY —That one marker per horizontal graticule line
is displayed.
n.
C H E C K -T h a t the spacing between the tim e markers
nearest the third and ninth vertical graticule lines is 6
divisions, ± 0 .0 6 division.
O.
Set HORIZONTAL M O D E to A.
p.
V ER IFY —That one marker per horizontal division
is displayed.
q.
C H E C K -T h a t the spacing between the tim e markers
nearest the third and ninth vertical graticule lines is
6 divisions, ± 0.06 division.
r.
Set HORIZONTAL M O D E to B.
s.
Repeat parts j through q to verify al IA and B acquisition
rate settings down to 50 0 ms.
t.
Disconnect the test setup.
3. Verify the DELAY TIME and A DELAY TIME
Functions, Check
A
DELAY TIME
Resolution, and Check Accuracy of the
Time-Base Reference (using the ADELAY
TIME function).
a.
Recall the Initial Front-Panel Setup, labeled “FPNL”
(see step i in “INITIAL S E T U P ” at the start of this
procedure). Make the following changes to the front-
panel setup:
Select: VERTICAL M O DE
Set:
CH 2
Off
Set:
CH 1 VOLTS/DIV
500 mV
A S EC /D IV
20 (AS
HORIZONTAL M O DE
A INTEN
B SEC /D IV
500 ns
A /B T R IG
B
b.
Select TRIGGER M O D E and set RUNS AFTER on. Set
A /B T R IG to A .
Perform ance Check and Functional Verification Procedure
2440 Service
c.
Use the HORIZONTAL POSITION control to align the
Trigger Point Indicator (a small “T ” on the displayed
trace) to the vertical graticule line 3 divisions left of
center screen.
d.
Connect the MARKER O U T signal of a Time Mark
Generator to the CH 1 OR X input connector through a
50 H Coaxial Cable. Do not use a Termination.
e.
Set the Tim e Mark Generator’s marker period to 20 p.s.
Vertically position the bottom of the CH 1 display to
1 division below center screen.
f.
Select DELAY TIM E and use the CURSOR/DELAY
control to adjust the DELAY T IM E Readout for a reading
of 120.00 |xs.
g.
V ER IFY —That the intensified zone is on the time
marker that is 3 divisions right of center screen.
h.
S e tth e HORIZONTAL M O DE to B. V ER IFY—the B Trig
ger Point Indicator is on the rising edge of the displayed
tim e marker.
i.
Set the HORIZONTAL M O DE to A INTEN and use the
HORIZONTAL POSITION control to position the A
Trigger Point Indicator to the graticule line 4 divisions
left of center screen.
j.
Use the CURSOR/DELAY control to adjust the DELAY
TIM E Readout for a reading of 20.00 |is (the intensified
zone will be aligned to the time marker 3 divisions left of
center screen).
k.
Press the A TIM E ON | OFF menu button to set
A TIM E ON.
l
.
Using the CURSOR/DELAY control,
adjust the
A DELAY TIM E Readout for a reading of 120.00 jis.
m. V E R IF Y -T h a t the A DELAY intensified zone is on the
marker 3 divisions right of center screen.
n.
Slightly rotate the CURSOR/DELAY control to increase
the A DELAY TIM E reading the least amount possible.
o.
C H E C K -T h a t the readout can be advanced in incre
ments at least as small as 0.04 p.s.
4-25
Summary of Contents for 2440
Page 4: ......
Page 12: ......
Page 38: ......
Page 91: ...Theory of Operation 2440 Service 3 49 Figure 3 5 Simplified CCD architecture ...
Page 120: ...Theory of Operation 2440 Service Figure 3 10 DC Restorer 3 78 ...
Page 130: ...Theory of Operation 2440 Service 3 88 ...
Page 136: ...Theory of Operation 2440 Service Figure 3 15 PWM Regulator and Inverter 3 94 ...
Page 138: ...Theory of Operation 2440 Service Figure 3 16 PWM switching waveforms 3 96 ...
Page 208: ......
Page 322: ......
Page 419: ......
Page 423: ...2440 Service Figure 9 4b 2440 Block diagram part 2 6330 27 BLOCK DIAGRAM PART 2 ...
Page 427: ......
Page 428: ...2440 Service WAVEFORMS FOR DIAGRAM 1 6 2 8 5 3 2 ...
Page 429: ......
Page 432: ...2440 Service WAVEFORMS FOR DIAGRAM 2 BUS ISOLATED 6 6 0 3 4 9 ...
Page 433: ...WAVEFORMS FOR DIAGRAM 2 ...
Page 439: ......
Page 445: ......
Page 450: ...2440 Service WAVEFORMS FOR DIAGRAM 8 6603 27 ...
Page 455: ...W A V EFO R M S FOR D IAG RAM 11 WAVEFORMS FOR DIAGRAM 11 6603 29 ...
Page 458: ...WAVEFORMS FOR DIAGRAM 12 ...
Page 463: ...WAVEFORMS FOR DIAGRAM 14 ...
Page 471: ... C N i u i t e x r 4 j u J l i 1 i 6 28 5 44 WAVEFORMS FOR DIAGRAM 18 ...
Page 476: ...WAVEFORMS FOR DIAGRAM 20 ...
Page 490: ...YES ...
Page 492: ......
Page 494: ......
Page 499: ...YES ...
Page 501: ......
Page 502: ...2440 Service FIG 1 C A B IN E T ...
Page 503: ......
Page 510: ...2440 Service SISSVHO Z Old ...
Page 511: ......
Page 512: ...A12 2440 Service F IG 3 C IR C U IT BOARDS ...
Page 513: ......
Page 517: ...10 16 R E V J U L 1993 Replaceable Mechanical Parts ...
Page 518: ...2440 Service ...
Page 519: ...FIG 4 L V H V PWR SUPPLY ...
Page 520: ...FIG 5 ACCESSORIES 2440 Service ...
Page 521: ......
Page 523: ......
Page 529: ......
Page 530: ......
Page 531: ......