![Tektronix 2440 User Manual Download Page 166](http://html1.mh-extra.com/html/tektronix/2440/2440_user-manual_1078860166.webp)
Performance Check and Functional Verification Procedure
2440 Service
HORIZONTAL SYSTEM
Equipment Required (see table 4-1)
Time-Mark Generator (Item 4)
50 f t Precision Coaxial Cable (Item 11)
50 f t Coaxial Cable (Qty 2) (Item 10)
50 f t Termination (Item 12)
1. Check Cursor Readout Accuracies for the A
and B Acquisition Systems.
a.
Recall the Initial Front-Panel Setup, labeled “FPNL”
(see step i in “INITIAL S E T U P ” at the start of this
procedure). Make the following changes to the front-
panel setup:
Set:
C H 1 VOLTS/DIV
1 V
Set:
CH 2 VOLTS/DIV
2 V
Select:
CURSOR FUNCTION
Set:
TIM E
On
b.
Use the CURSOR/DELAY control to align the movable
cursor (it will have more dots than the alternate cursor)
to the third graticule line to the left of center screen.
c.
Press CURSOR SELECT to enable the alternate cursor.
d.
Use the CURSOR/DELAY control to align cursor to the
third graticule line to the right of center screen.
e.
C H E C K -T h a t the Cursor Time Readout indicates
2.9700 to 3.0300 ms.
f.
Set the HORIZONTAL MODE to B.
g.
C H E C K -T h a t the Cursor Time Readout indicates
2.9700 to 3.0300 ms.
2. Verify the Sample Rate of the A and B
Acquisition Systems and Check the
Horizontal Display Accuracy.
a.
Recall the Initial Front-Panel Setup, labeled “FPNL”
(see step i in “INITIAL S ETU P ” at the start of this
procedure). Make the following changes to the front-
panel setup:
Select: VERTICAL M O DE
Set:
CH 2
Off
Set:
CH 1 VOLTS/DIV
500 mV
A S E C /D IV
100 ns
A /B T R IG
B
b.
Select TRIGG ER M O D E and set RUNS AFTER on. Set
A/B TRIG to A.
c.
Connect the MARKER O U T signal of a Time Mark
Generator to the CH 1 OR X input connector through a
50 f t Coaxial Cable. Do not use a Termination.
d.
Set the Time Mark Generator’s marker period to 0.1 p.s.
e.
Push IN IT @ 50% to set the A Trigger level.
f.
V E R IF Y -T h a t one time marker per horizontal division
is displayed.
g.
C H E C K —That the spacing between the time markers
nearest the third and ninth vertical graticule lines is 6
divisions, ± 0 .0 6 division.
h.
Set HORIZONTAL M O DE to B and set the B SEC /D IV
control to 100 ns.
i.
V E R IF Y —That one marker per horizontal division
is displayed.
j.
C H E C K -T h a t the spacing between the time markers
nearest the third and ninth vertical graticule lines is
6 divisions, ± 0 .0 6 division.
k.
Rotate the A and B S EC /D IV control counter-clockwise
one position to set both acquisition systems one
speed slower.
4-24
Summary of Contents for 2440
Page 4: ......
Page 12: ......
Page 38: ......
Page 91: ...Theory of Operation 2440 Service 3 49 Figure 3 5 Simplified CCD architecture ...
Page 120: ...Theory of Operation 2440 Service Figure 3 10 DC Restorer 3 78 ...
Page 130: ...Theory of Operation 2440 Service 3 88 ...
Page 136: ...Theory of Operation 2440 Service Figure 3 15 PWM Regulator and Inverter 3 94 ...
Page 138: ...Theory of Operation 2440 Service Figure 3 16 PWM switching waveforms 3 96 ...
Page 208: ......
Page 322: ......
Page 419: ......
Page 423: ...2440 Service Figure 9 4b 2440 Block diagram part 2 6330 27 BLOCK DIAGRAM PART 2 ...
Page 427: ......
Page 428: ...2440 Service WAVEFORMS FOR DIAGRAM 1 6 2 8 5 3 2 ...
Page 429: ......
Page 432: ...2440 Service WAVEFORMS FOR DIAGRAM 2 BUS ISOLATED 6 6 0 3 4 9 ...
Page 433: ...WAVEFORMS FOR DIAGRAM 2 ...
Page 439: ......
Page 445: ......
Page 450: ...2440 Service WAVEFORMS FOR DIAGRAM 8 6603 27 ...
Page 455: ...W A V EFO R M S FOR D IAG RAM 11 WAVEFORMS FOR DIAGRAM 11 6603 29 ...
Page 458: ...WAVEFORMS FOR DIAGRAM 12 ...
Page 463: ...WAVEFORMS FOR DIAGRAM 14 ...
Page 471: ... C N i u i t e x r 4 j u J l i 1 i 6 28 5 44 WAVEFORMS FOR DIAGRAM 18 ...
Page 476: ...WAVEFORMS FOR DIAGRAM 20 ...
Page 490: ...YES ...
Page 492: ......
Page 494: ......
Page 499: ...YES ...
Page 501: ......
Page 502: ...2440 Service FIG 1 C A B IN E T ...
Page 503: ......
Page 510: ...2440 Service SISSVHO Z Old ...
Page 511: ......
Page 512: ...A12 2440 Service F IG 3 C IR C U IT BOARDS ...
Page 513: ......
Page 517: ...10 16 R E V J U L 1993 Replaceable Mechanical Parts ...
Page 518: ...2440 Service ...
Page 519: ...FIG 4 L V H V PWR SUPPLY ...
Page 520: ...FIG 5 ACCESSORIES 2440 Service ...
Page 521: ......
Page 523: ......
Page 529: ......
Page 530: ......
Page 531: ......