Safety
Information about Personal Safety
0.0
04/2012
0.0
A.2-27
0.0
Calibration
A.2
If the CT system is not properly setup (for example, no check-up
performed) CT scale displacements and inhomogeneities can
result. This may result just after switching on the unit, during
warm-up to operational temperature (calibration), or due to an
extended length of service of the X-ray tube. The same applies
to defects in the measurement system.
A.2
The daily quality checkup usually detects and reports such dis-
crepancies.
A.2
B
.2
A
Additional calibrations may also be necessary if large changes in
temperature occur in the room.
A.2
Other system defects or calibration deviations cause streaks or
(partial) rings to appear in the CT image.
A.2
Summary of Contents for SOMATOM Spirit
Page 4: ...0 0 0 0 ii SOMATOM Spirit0 0 ...
Page 8: ...0 0 0 0 vi SOMATOM Spirit0 0 ...
Page 55: ...Safety Information about Personal Safety 0 0 04 2012 0 0 A 2 130 0 A 2 ...
Page 72: ...0 0 0 0 Information about Personal Safety Safety A 2 30 SOMATOM Spirit0 0 ...
Page 84: ...0 0 0 0 Mechanical Safety Safety A 3 12 SOMATOM Spirit0 0 ...
Page 128: ...0 0 0 0 Software Related Safety Information Safety A 4 44 SOMATOM Spirit0 0 ...
Page 164: ...0 0 0 0 Console Related Safety Advice System Description B 4 8 SOMATOM Spirit0 0 ...
Page 190: ...0 0 0 0 Storage Media and Drives System Description B 6 6 SOMATOM Spirit0 0 ...
Page 198: ...0 0 0 0 System Handling C 2 SOMATOM Spirit0 0 ...
Page 230: ...0 0 0 0 Use of Storage Media System Handling C 3 14 SOMATOM Spirit0 0 ...
Page 234: ...0 0 0 0 Operation D 4 SOMATOM Spirit0 0 ...
Page 266: ...0 0 0 0 Function Tests Operation D 2 6 SOMATOM Spirit0 0 ...
Page 350: ...0 0 0 0 Maintenance E 2 SOMATOM Spirit0 0 ...
Page 362: ...0 0 0 0 Cleaning Agents and Disinfectants Maintenance E 2 6 SOMATOM Spirit0 0 ...
Page 366: ...0 0 0 0 Cleaning and Disinfecting Maintenance E 3 4 SOMATOM Spirit0 0 ...
Page 368: ...0 0 0 0 Quality Assurance F 2 SOMATOM Spirit0 0 ...
Page 376: ...0 0 0 0 General Information Quality Assurance F 1 8 SOMATOM Spirit0 0 ...
Page 384: ...0 0 0 0 Daily Quality Measurements Quality Assurance F 2 8 SOMATOM Spirit0 0 ...
Page 424: ...0 0 0 0 Constancy Test Quality Assurance F 3 40 SOMATOM Spirit0 0 ...
Page 436: ...0 0 0 0 Low Contrast Test Optional Quality Assurance F 4 12 SOMATOM Spirit0 0 ...
Page 439: ...Quality Assurance Camera Test 0 0 04 2012 0 0 F 5 30 0 Examples of test images F 5 F 5 ...
Page 450: ...0 0 0 0 Technical Specifications G 2 SOMATOM Spirit0 0 ...
Page 452: ...0 0 0 0 Standards and Statutory Regulations H 2 SOMATOM Spirit0 0 ...
Page 454: ...0 0 0 0 Forms J 2 SOMATOM Spirit0 0 ...
Page 456: ...0 0 0 0 DIP Report K 2 SOMATOM Spirit0 0 ...
Page 458: ...0 0 0 0 Location of Labels L 2 SOMATOM Spirit0 0 ...