R&S
FS-K30
Measurements & Settings
Measurement
Settings
Software Manual 1300.6637.42 - 03
67
2.5.4.5 Init before Meas
Init Before Meas
is used to specify whether the external signal generator is initialised
before each measurement or not.
When
Init Before Meas
is set to On, an initialisation sequence of GPIB commands will be
sent to the signal generator before each measurement. This ensures that the signal
generator will be in the correct state to receive control commands during measurement
execution. However, this will add a time overhead to the overall measurement completion.
When
Init Before Meas
is set to Off, no initialisation sequence of GPIB commands will
be sent to the signal generator when a measurement is run.
Note: The INIT GEN softkey in the main measurement results view and measurements
settings view can be pressed to get R&S FS-K30 to send the initialisation sequence of
GPIB commands to the signal generator at any time.
This parameter is only available when
Automatic
Control
is set to ON. This parameter
is not displayed when the B10 option is not installed.
2.5.4.6 Auto Switch OFF
When the Auto Switch Off option is selected R&S FS-K30 will automatically switch the
external generator's RF output off after a measurement completes or is aborted.
This parameter is only available only when Automatic Control is selected.
2.5.5 Generator Frequency
When an external signal generator (or VCO) is used to create a Local Oscillator (
LO
)
frequency, the following formula is used to calculate the generator frequency:
Generator Frequency = [(
LO
+
Offset 1
) *
Factor 1
/
Factor 2
)] +
Offset 2
The fields in the calculation can be edited directly. As the calculation is updated the
minimum and maximum generator frequencies required based on the values in the
calculation and the minimum and maximum LO frequencies are calculated and displayed.