3.1.10 INPut Subsystem
The INPut subsystem controls the input characteristics of the RF inputs of the
instrument.
INPut:ATTenuation <numeric_value>
This remote control command specifies the RF attenuation that the analyser
imposes.
Example
INP:ATT 10
'attenuation of the analyzer is set to 10 dB.
Characteristics
*RST value: 0 dB
SCPI: conforming
INPut:PRESelection[:STATe] ON | OFF
This remote control command switches the preselection on or off.
Example
INP:PRES:STAT ON
'preselection is switched on.
Characteristics
*RST value: OFF
SCPI: device-specific
The command is only available with the preselector option B2.
INPut:GAIN:STATe ON | OFF
This remote control command switches the pre-amplifier on and off. If only the B2
or B25 option is present then the pre-amplifier of the installed option is controlled. If
both the B2 and B25 options are present then the pre-amplifier controlled by this
command depends on the setting for the INPut:PRESelection[:STATe] command.
If the INPut:PRESelection[:STATe] is set to ON then the pre-amplifier for the B2
option is controlled by this command, otherwise the pre-amplifier of the B25 option
is controlled.
Example
INP:GAIN:STAT ON
'switches the pre-amplifier
Characteristics
*RST value: OFF
SCPI: device-specific