R&S
FS-K30
Remote Control
Description
of commands
Software Manual 1300.6637.42 - 03
97
CALCulate:LIMit<1 to 6>:UPPer[:DATA] <numeric_value>,<numeric_value>...
This command defines the values for the upper limit lines independently of the
measurement window.
The number of values for the CONTrol axis and for the corresponding UPPer
and/or LOWer limit line have to be identical. Otherwise default values are entered
for missing values or unnecessary values are deleted.
The unit must be identical with the unit selected by
CALC:LIM:UNIT
. If no unit is
indicated, the unit defined with
CALC:LIM:UNIT
is automatically used.
Example:
CALC:LIM2:UPP -10,0,0,-10,-5
'Defines 5 upper limit values for limit line 2 in the preset
unit.
CALC:LIM2:UPP?
'Outputs the upper limit values for limit line 2 separated by
a comma.
Characteristics
*RST value: -
SCPI: conforming
CALCulate:LIMit<1 to 6>:UPPer:STATe ON | OFF
This command switches on or off the indicated limit line. The limit check is
activated separately with
CALC:LIM:STAT ON
Example
CALC:LIM4:UPP:STAT ON
'Switches on limit line 4 (upper limit)
Characteristics
*RST value: -
SCPI: conforming
CALCulate:LIMit<1 to 6>:UPPer:SHIFt <numeric_value>
This command moves a limit line by the indicated value in Y-direction. The line is
shifted by modifying the individual y values
Example
“
CALC:LIM3:UPP:SHIF 20DB
'Shifts all Y values of limit line 3 by 20 dB.
Characteristics
*RST value: -
SCPI: device-specific
This command is an event and is therefore not assigned an *RST value and has
no query.