Example
DISP:TRAC:Y:AUTO ON
'Switches on automatic scaling of the Y-axis for all traces
Characteristics
*RST value: ON for both TRACe1 and TRACe2
SCPI: conforming
DISPlay[:WINDow<1>]:TRACe<1|2>:Y[:SCALe]:BOTTom <numeric_value>
This command sets the minimum (bottom) Y-axis display value for the specified
trace display. Note that this command has no affect if automatic scaling of the Y-
axis is enabled or the specified trace is not currently active.
Example
DISP:TRAC:Y:BOTT –30
'Sets the minimum Y-axis display to –30 dB for trace 1
Characteristics
*RST value: 0 dBm
SCPI: conforming
DISPlay[:WINDow<1>]:TRACe<1|2>:Y[:SCALe]:RLEVel <numeric_value>
This command sets the reference level for the Y-axis display value for all trace
displays.
Example
“
DISP:TRAC:Y:RLEV 0
”
Sets the reference level 0 dBm
Characteristics
*RST value: -30 dBm
SCPI: conforming
DISPlay[:WINDow<1>]:TRACe<1|2>:Y[:SCALe]:RLEVel:AUTO ON | OFF
This command enables or disables automatic reference level detection.
Example
DISP:TRAC:Y:RLEV:AUTO ON
'turns on automatic reference level detection.
Characteristics
*RST value: ON