R&S
FS-K30
Measurements & Settings
Measurement
Settings
Software Manual 1300.6637.42 - 03
61
at low frequencies, the
RBW
must be reduced to prevent the LO frequency of the analyser
from invalidating the measurement. At receive frequencies of 100 kHz, the
RBW
must not
exceed 10 kHz.
2.5.3.2 Sweep Time
The
Sweep Time
is the time taken by the spectrum analyzer to perform one complete
measurement sweep (measurement step). Note that two sweeps are performed for
each measurement step (once for noise source on, once with noise source off).
For narrow bandwidths, the
Sweep Time
should be increased in order to give accurate
measurement results
2.5.3.3 Settling Time
The
Settling Time
is the time taken for the DUT to settle after a noise source has been
turned on or off.
Most noise sources generate an interfering DC component in addition to the noise
spectrum. When the noise source is switched on or off, low-frequency DUTs may
require this
Settling Time
for coupling capacitors to be charged or discharged.