General Information
Quick
Start Guide
Software Manual 1300.6637.42 - 03
13
The number of steps can be increased to up to 100 frequency points. This would,
of course, result in a correspondingly longer measurement time.
5. Press the
ENR
softkey to open the ENR Settings view
The default for the ENR value is 15 dB. The ENR value can either be entered as a
constant value that is valid for all frequencies (
ENR Constant
) or as a list of
frequency-dependent ENR values in the table on the right-hand-side of the ENR
Settings view.
For a noise source with a frequency-dependent ENR, sampled ENR values must
be entered for a number of different frequencies. The manufacturer of the noise
source supplies these sampled values.
For the purposes of this introduction to R&S FS-K30, it is sufficient to specify a
constant ENR value for this measurement. The default
Selection
for ENR is a
Constant value for all frequencies, so this does not need to be changed.
Enter the ENR value that is valid for the measurement range of 220 MHz to 320
MHz in the
ENR Constant
field.
Selection
ENR Settings
Constant
ENR Constant
15 dB
Room Temperature
290 K