![Rohde & Schwarz FS-K30 Software Manual Download Page 121](http://html.mh-extra.com/html/rohde-and-schwarz/fs-k30/fs-k30_software-manual_1477829121.webp)
R&S
FS-K30
Remote Control
Description
of commands
Software Manual 1300.6637.42 - 03
105
DISPlay[:WINDow<1>]:TRACe<1|2>:Y[:SCALe]:TOP <numeric_value>
This command sets the maximum (top) Y-axis display value for the specified trace
display. Note that this command has no affect if automatic scaling of the Y-axis is
enabled or the specified trace is not currently active.
Example
“DISP:TRAC:Y:TOP 30
'Sets the maximum Y-axis display to 30 dB for trace 1
Characteristics
*RST value: 20 dB
SCPI: conforming
DISPlay[:WINDow<1>]:TABLe
This command toggles the display of results in graphical or tabular form.
Example
“
DISP:WIND1:TABL ON
“
The R&S FS-K30 option will display the table of results
Characteristics
*RST value: OFF
SCPI: device-specific
3.1.7 FETCh Subsystem
The FETCh subsystem retrieves results for the most recently completed fixed
frequency or frequency list measurements. Frequency list results are returned as a list
of results where the result is that requested in the specific fetch command. Single
frequency results are single numbers in the described units.
Corrected measurements are only accessible after a user calibration has been
performed.
FETCh:ARRay:MEMory<1to3>:NOISe:FIGure?
FETCh:ARRay:MEMory<1to3>:NOISe:TEMPerature?
FETCh:ARRay:MEMory<1to3>:NOISe:GAIN?
These remote control commands request the FS-K30 option to return the noise
measurement results in the selected memory. The results will be returned as a 100
element array, either of noise figure, noise temperature or noise gain results.