2 Measurements & Settings
This section contains a detailed description of all measurement modes, settings &
results. It covers the following subjects:
I
Measurement modes & schematics
I
Calibration
I
Running measurements
I
Measurement results
I
Frequency settings
I
Measurement settings (including control of External Signal Generator)
I
ENR settings
I
Loss settings
I
Graphic settings
I
General hints about noise measurements
2.1 Measurement modes & schematics
Noise measurements are performed on many different types of Device Under Test
(DUT). The type of DUT to be measured determines the test set-up and also how the
frequency list is to be generated. To support these different types of DUT, R&S FS-K30
provides five different measurement modes:
I
Direct
I
Fixed IF, LO = RF + IF
I
Fixed IF, LO = abs(RF – IF)
I
Fixed LO, IF = RF + LO
I
Fixed LO, IF = abs(RF – LO)
The
Measurement Mode
can be selected in the Frequency Settings view by pressing
the
SET FREQ
softkey.
R&S FS-K30 also provides a schematic display of the test set-up for each of the above
measurement modes. This provides a diagrammatic summary of the test set-up and
frequency ranges. It cannot be used to enter or change any settings. The schematic
display can be viewed by pressing the
SCHEMATIC
softkey within the Frequency
Settings view. The upper part of the display shows the set-up for calibration and the
lower part of the display shows the test set-up for the measurement proper.
2.1.1 Direct measurements
The Direct setting should be used for DUTs without frequency-conversion, for example,
amplifiers. In such cases, a local oscillator is not required and the frequency being
measured is the RF frequency.
The schematic display for a Direct measurement mode is shown in Fig. 22.