R&S
FS-K30
Remote Control
Description
of commands
Software Manual 1300.6637.42 - 03
111
[SENSe]:CORRection:ENR[:MEASurement]:TABLe:DATA
<numeric_value>,<numeric_value>,...
This ENR remote control command allows an ENR lookup table list to be specified
that will be used by the K30 option to determine the correct ENR (excess noise
ratio) figure to use for the input frequency that requires measuring. The list
arguments specified will completely overwrite all current ENR frequency list entries
regardless of how many entries are present and how many entries are being supplied
for the new list.
This command will however not be effective if the remote control command
SENS:CORR:ENR:MODE SPOT was previously entered.
numeric value :=
a frequency ENR pair of arguments, up to a maximum of
argument 100 pairs. The frequency can be specified in Hz, kHz, MHz or GHz. The
ENR figure must be a value in dB.
Example
SENS:CORR:ENR:MEAS:TABL:DATA 1MHZ,10,2MHZ,12
'The R&S FS-K30 option will overwrite the current ENR list
with the two entry pairs specified.
Characteristics
*RST value: -
SCPI: device specific
[SENSe]:CORRection:TEMPerature <numeric_value>
This ENR remote control command allows the room temperature of the operating
environment to be specified. This value will be taken into account when calculating
noise results.
Example
SENS:CORR:TEMP 291.50
'The R&S FS-K30 option will factor in a room temperature of
291.50 Kelvin (18.5 C) when performing noise measurements.
Characteristics
*RST value: 293 K
SCPI: device specific
[SENSe]:CORRection:LOSS:INPut:MODE SPOT | TABLe
This remote control command is used to identify whether a single input loss value
applies for all measured frequencies or whether an input loss list is to be used. (Loss
Input values specified at specific input frequencies).
Parameter
SPOT = The Constant loss input value for all measurement frequencies is used.
TABLe = The loss input table is used.