![Rohde & Schwarz FS-K30 Software Manual Download Page 110](http://html.mh-extra.com/html/rohde-and-schwarz/fs-k30/fs-k30_software-manual_1477829110.webp)
R&S
FS-K30
Remote Control
Description
of commands
Software Manual 1300.6637.42 - 03
94
Example:
CALC:LIM1:COPY 2
'Copies limit line 1 to line 2.
CALC:LIM1:COPY 'NFIG2'
'Copies limit line 1 to a new line named 'NFIG2'.
Characteristics
*RST value: -
SCPI: device-specific
This command is an event and is therefore not assigned an *RST value and has
no query.
CALCulate:LIMit<1 to 6>:NAME <name>
This command assigns a name to a limit line numbered 1 to 6. If it does not exist
already, a limit line with this name is created.
name = name of the new limit line given as a string
Example:
CALC:LIM1:NAME 'NFIG1'
'Assigns the name 'NFIG1' to limit line 1.
Characteristics
*RST value: 'REM1' to 'REM8' for lines 1 to 8
SCPI: device-specific
CALCulate:LIMit<1 to 6>:DELete
This command deletes the selected limit line.
Example:
CALC:LIM1:DEL
'Deletes limit line 1.
Characteristics
*RST value: -
SCPI: device-specific
This command is an event and is therefore not assigned an *RST value and has
no query.
3.1.3.2 CALCulate:LIMit:CONTrol Subsystem
The CALCulate:LIMit:CONTrol subsystem defines the x-axis (CONTrol-axis).