R&S
FS-K30
Remote Control
Description
of commands
Software Manual 1300.6637.42 - 03
116
[SENSe]:FREQuency:STARt <numeric value>
This remote control command can be used to specify the starting frequency for a
new frequency measurement list that requires computing. This value will form the
basis of one of the input criteria for computing a new frequency list.
Example
SENS:FREQ:STARt 500MHZ
'the starting frequency for a new list, yet to be created, is
set to a value of 500 MHz.
Characteristics
*RST value: 550 MHz
SCPI: conforming
[SENSe]:FREQuency:STOP <numeric value>
This remote control command can be used to specify the ending frequency for a
new frequency measurement list that requires computing. This value will form the
basis of one of the input criteria for computing a new frequency list.
Example
SENS:FREQ:STOP 700MHZ
'the ending frequency for a new list, yet to be created, is
set to a value of 700 MHz.
Characteristics
*RST value: 560 MHz
SCPI: conforming
[SENSe]:FREQuency:STEP <numeric value>
This remote control command can be used to specify the step frequency for a new
frequency measurement list that requires computing. This value will form the basis
of one of the input criteria for computing a new frequency list.
Example
SENS:FREQ:STEP 10MHZ
the step frequency for a new list, yet to be created, is set
to a value of 10 MHz.
Characteristics
*RST value: 2 MHz
SCPI: device-specific
[SENSe]:FREQuency[:CW|:FIXed] <numeric value>
This remote control command is used to specify a single fixed frequency to
measure noise and gain continuously. The value specified with this command is
only used when a single frequency measurement is initiated.