R&S
FS-K30
Remote Control
Description
of commands
Software Manual 1300.6637.42 - 03
115
Characteristics
*RST value: OFF
SCPI: device specific
3.1.12.2 Measurement Settings
The following diagram shows the Measurement Settings SENSe subsystem:
[SENSe]:BANDwidth|BWIDth:RESolution <numeric value>
This remote control allows the resolution bandwidth of the analyser to be specified.
Example
SENS:BAND:RES 1MHZ
'The R&S FS-K30 option will set the resolution bandwidth to1
MHz.
Characteristics
*RST value: 1 MHz
SCPI: conforming
[SENSe]:SWEep:TIME <numeric value>
This remote control command allows the sweep time to be specified.
Example
SENS:SWE:TIME 100MS
'sets sweep time to 100 milliseconds.
Characteristics
*RST value: 100 milliseconds
SCPI: conforming
[SENSe]:SWEep:COUNt <numeric value>
This remote control specifies the number of sweeps over which the measurement
results are to be averaged. The higher the number of sweeps averaged the more
accurate the results however the more sweeps required the longer the
measurement takes to perform.
Example
SENS:SWE:COUN 10
'sets the averaging of the measurement to be performed over
10 sweeps
Characteristics
*RST value: 1
SCPI: conforming