R&S
FS-K30
Measurements & Settings
Measurement
Settings
Software Manual 1300.6637.42 - 03
66
the Fixed LO cases in which the signal generator settings do not need to change
during the measurement. In the Fixed IF
Measurement Modes
, meaningful results will
only be possible if
Automatic Control
is set to ON, since R&S FS-K30 will expect the
LO
frequency to be automatically tuned to the test
RF
frequency.
When
Automatic Control
is set to ON, R&S FS-K30 will issue GPIB commands during
measurement execution in order to control the signal generator.
Note that option B10 must be installed on the spectrum analyzer for GPIB control of an
external signal generator to be possible. When B10 is not installed, the
Automatic
Control
parameter will be set to OFF and will not be selectable.
2.5.4.2 Level
Level
is the desired output power level of the external signal generator.
This parameter is only available when
Automatic
Control
is set to ON. This parameter
is not displayed when the B10 option is not installed
2.5.4.3 Generator Select
Generator Select
is used to select an external signal generator model from the defined list.
This is necessary to ensure that the correct GPIB commands are sent to the signal generator.
The list of signal generators is defined within the spectrum analyzer software – refer to
the spectrum analyzer and B10 option user manuals for details on how to edit this list
and add new signal generators.
This parameter is only available when
Automatic
Control
is set to ON. This parameter
is not displayed when the B10 option is not installed
2.5.4.4 GPIB Address
GPIB Address
is the address used to control the external signal generator via GPIB.
This parameter is only available when
Automatic
Control
is set to ON. This parameter
is not displayed when the B10 option is not installed