![Rohde & Schwarz FS-K30 Software Manual Download Page 145](http://html.mh-extra.com/html/rohde-and-schwarz/fs-k30/fs-k30_software-manual_1477829145.webp)
R&S
FS-K30
Remote Control
Description
of commands
Software Manual 1300.6637.42 - 03
129
Example
SYST:CONF:GEN:CONT:STAT ON
'The R&S FS-K30 option completely controls the setting of the
external signal generator.
Characteristics:
*RST value: OFF
SCPI: device-specific
SYSTem:CONFigure:GENerator:INITialise:AUTO ON | OFF
This remote control command is used to specify whether an initialisation sequence
of GPIB commands is sent to an external signal generator prior to performing each
measurement. Sending an initialisation sequence before each measurement
ensures that the signal generator will be in the correct state to receive control
commands during a measurement sequence, however initialising an external
signal generator before each measurement adds a time overhead to each
measurement.
Example
“SYST:CONF:GEN:INIT:AUT
O
ON
'The R&S FS-K30 option initialises the selected signal
generator prior to starting each requested measurement.
Characteristics:
*RST value: OFF
SCPI: device-specific
SYSTem:CONFigure: GENerator:SWITch:AUTO ON | OFF
This remote control command is used to specify whether a sequence of GPIB
commands is sent to an external signal generator at the end of each measurement
(after a single sweep or after a measurement is aborted) to switch the generator’s
RF output off.
Example:
SYST:CONF:GEN:SWIT:AUT
O
ON
'The FS-K30 option switches the generator’s RF output after
each measurement.
Characteristics:
*RST value: OFF
SCPI: device-specific
SYSTem:CONFigure:DUT:GAIN <numeric value> <numeric value>
This remote control command allows the maximum gain of the DUT to be
specified.