CALCulate:LIMit<1 to 6>:CONTrol[:DATA] <numeric_value>,<numeric_value>
This command defines the X-axis values (frequencies) of the upper or lower limit lines.
The number of values for the CONTrol axis and for the corresponding UPPer
and/or LOWer limit lines have to be identical. Otherwise default values are entered
for missing values or unnecessary values are deleted.
Example:
CALC:LIM2:CONT 1MHz,30MHz,100MHz, 300MHz,1GHz
'Defines 5 reference values for the X-axis of limit line 2
CALC:LIM2:CONT?
'Outputs the reference values for the X-axis of limit line 2
separated by a comma.
Characteristics
*RST value: -
SCPI: conforming
CALCulate:LIMit<1 to 6>:CONTrol:SHIFt <numeric_value>
This command moves a limit line by the indicated value in x direction. The line is
shifted by modifying the individual x values.
Example:
CALC:LIM2:CONT:SHIF 50KHZ
'Shifts all reference values of limit line 2 by 50 kHz.
Characteristics
*RST value: -
SCPI: device-specific
This command is an event and is therefore not assigned an *RST value and has
no query.
3.1.3.3 CALCulate:LIMit:LOWer Subsystem
The CALCulate:LIMit:LOWer subsystem defines the lower limit line. Note that if a set
command is issued in this subsystem the limit line effected is automatically converted
to a lower limit line.
CALCulate:LIMit<1 to 6>:LOWer[:DATA] <numeric_value>,<numeric_value>...
This command defines the values for the selected lower limit line.
The number of values for the CONTrol axis and for the corresponding LOWer limit
line has to be identical. Otherwise default values are entered for missing values or
unnecessary values are deleted.
If the measured values are smaller than the LOWer limit line, the limit check
signals errors.