R&S
FS-K30
Remote Control
Description
of commands
Software Manual 1300.6637.42 - 03
92
CALCulate:LIMit<1 to 6>:TRACe NFIGure | TEFFective | GAIN
This command assigns a limit line to a particular measurement type. Measurement
types include noise figure, noise temperature or gain.
Parameter
NFIGure
= Noise Figure results
TEFFective = Noise Temperature results
GAIN
= Gain results
Examples
CALC:LIM2:TRAC NFIG
'Assigns limit line 2 to the noise figure measurement.
CALC:LIM3:TRAC GAIN
'Assigns limit line 3 to trace the gain measurement.
Characteristics
*RST value: -
SCPI: device-specific
CALCulate:LIMit<1 to 6>:STATe ON | OFF
This command switches on or off the limit check for the selected limit.
The result of the limit check can be queried with
CALCulate:LIMit<1 to
6>:FAIL?
.
Example
CALC:LIM:STAT ON
'Switches on the limit check for limit line 1
CALC:LIM:STAT OFF
'Switches off the limit check for limit line 1
Characteristics
*RST value: -
SCPI: conforming
CALCulate:LIMit<1 to 6>:FAIL?
This command queries the result of the limit check of the indicated limit line. It
should be noted that a complete sweep must have been performed for obtaining a
valid result. A synchronization with *OPC, *OPC? or *WAI should therefore be
provided. The result of the limit check responds with 0 for PASS and 1 for FAIL.