R&S
FS-K30
Remote Control
Description
of commands
Software Manual 1300.6637.42 - 03
128
Generator
Generator
Min Freq
Generator Max
Freq
Generat
or
Min
Power
dBm
Generat
or
Max
Power
dBm
SMP04
10 MHz
40 GHz
-130
3)
+12
3)
SMP22
10 MHz
20 GHz
-130
3)
+20
3)
SMU200A
100 kHz
2:2 GHz
-145
+13
SMU200B31
100 kHz
2:2 GHz
-145
+19
SMU03
100 kHz
3 GHz
-145
+13
SMU03B31
100 kHz
3 GHz
-145
+19
SMU04
100 kHz
4 GHz
-145
+13
SMU04B31
100 kHz
4 GHz
-145
+19
SMU06
100 kHz
2:2 GHz
-145
+13
SMU06B31
100 kHz
6 GHz
-145
+13
SMT02
5.0 kHz
1.5 GHz
-144
+13
SMT03
5.0 kHz
3.0 GHz
-144
+13
SMT06
5.0 kHz
6.0 GHz
-144
+13
SMX
100 kHz
1.0 GHz
-137
+13
SMY01
9 kHz
1.04 GHz
-140
+13
SMY02
9 kHz
2.08 GHz
-140
+13
HP8340A
10 MHz
26.5 GHz
-110
10
HP ESG-A Series 1000A,
2000A, 3000A, 4000A
250 kHz
4 GHz
-136
20
HP ESG-D SERIES
E4432B
250 kHz
3 GHz
-136
+10
1)
Requires mounting of option SMR-B11.
NOTICE!
Generators with TTL interface can also be operated via IECBUS (= GPIB) alone.
The command is only available with option Ext. Generator Control B10.
Example
SYST:COMM:RDEV:GEN:TYPE 'SME02'
'selects SME02 as the external signal generator
Characteristics:
*RST value: NONE
SCPI: device-specific
SYSTem:CONFigure:GENerator:CONTrol:STATe ON | OFF <boolean>
This remote control command is used to specify whether the setup of the external
generator is to be automatically controlled by the R&S FS-K30 option (via GPIB) or
manually by the user.