R&S
FS-K30
Remote Control
Status
reporting registers
Software Manual 1300.6637.42 - 03
134
STATus:QUEStionable:CORRection Register
This register comprises information about the correction state of noise measurements in R&S
FS-K30. It can be queried by commands
STATus:QUEStionable:CONDition?
and
STATus:QUEStionable[:EVENt]
?.
Bit No
Meaning
0
NO CORRection
User calibration is required (i.e. not done, or setup changed). Will remain 1 until a user calibration is done.
Set to 1 at the start of a user calibration. It will go to 0 at the end of a user calibration only if at least all
points on one range have been calibrated. Initial value is 1.
1
These bits are not used
2
UNCorrectected measurement
Uncorrected measurement data (one or more points could not be corrected using existing user
calibration). Set to 0 at the start of each sweep/redisplay of result. Will remain zero until an attempt is
made to correct a point and the calibration data does not exist (the required range has not been
calibrated). Note that if no user calibration data exists, this bit will not be set when an attempt is made to
make a corrected measurement — use Bit 0 to determine if a corrected measurement can be attempted.
3 to 14
These bits are not used
15
This bit is always 0
3.2.2 Error Reporting
Error reporting for the K30 option is carried out using the Service Request (SRQ)
interrupt in the GPIB interface. When an error occurs a Service Request interrupt will
be generated. The master can then query the slave instrument for the error that
triggered the interrupt Errors are queried through the “SYSTem:ERRor” command.