2.8.5 Noise and Gain X-Axis Settings
2.8.5.1 X-Axis
The X
-Axis
setting allows user to configure the x-axis to the RF or IF frequencies.
This setting is only available when the measurement mode is other than DIRECT.
2.8.6 Measurements in Detail
This section provides a more detailed explanation of the measurements provided by
R&S FS-K30 and provides help for using R&S FS-K30 to measure the characteristics
of specific types of DUT.
2.8.7 DUTs with Very Large Gain
If the gain of the DUT exceeds 60 dB, the total gain must be reduced by an external
attenuator. The total gain of the DUT together with the external attenuator should lie
within the range 10 dB to 60 dB. A total gain of 20 dB to 30 dB is recommended. For a
DUT with a gain of e.g. 64 dB, it is recommended to use an external 40-dB attenuator.
If an external attenuator is used the entry in the
Range
field in the Measurement
Settings view should be modified according to the total gain ( =
G
dut
– external
attenuator).
The attenuation values of the external attenuator are entered in the
LOSS OUTPUT
settings in the Loss Settings view.
Inaccuracies when entering this attenuation mainly influence the measured gain. The
noise figure remains to a large extent unaffected.
Low pass
Noise Source
0V
+28V
Noise 28V
RF
Low pass
Noise Source
DUT
0V
+28V
Noise 28V
RF
external
Attenuator
Fig. 42 Calibration and measurement on DUTs with a high gain