![Rohde & Schwarz FS-K30 Software Manual Download Page 66](http://html.mh-extra.com/html/rohde-and-schwarz/fs-k30/fs-k30_software-manual_1477829066.webp)
R&S
FS-K30
Measurements & Settings
Measurement
results
Software Manual 1300.6637.42 - 03
50
Fig. 26 A typical tabular display of measurement results
I
The tabular section below the title bar shows the overall measurement settings
used for the last measurement. This includes the following:
I
RBW: Resolution Bandwidth (Hz)
I
Average
I
RF Atten: RF Attenuation (dB)
I
Ref Meas: Manual if the reference level is manually entered, Automatic if the
reference level is measured automatically before each measurement
I
2
nd
Stage Corr.: On if 2
nd
stage correction is applied using calibration data, Off if no
correction is applied to the measurement results
I
Image Reject: image rejection of the DUT (dB)
The Current Value section of the table shows the settings and measurement results for
the currently selected measurement point in the frequency list. At the completion of a
measurement, this will display the settings & results of the last point in the list. When
another point in the frequency list is selected, the Current Value section will update
accordingly. This includes the following:
I
RF: Receive Frequency at the DUT at which the current values were measured (Hz)
I
LO: Local Oscillator frequency (Hz) – not displayed for Direct measurements
I
IF: Intermediate Frequency (Hz) – not displayed for Direct measurements
I
ENR: ENR value (dB) – refers to the Receive Frequency (RF)
I
Loss In: Loss at the Input of the DUT (dB) – refers to the Receive Frequency (RF)
I
Loss Out: Loss at the Output of the DUT (dB) – refers to the Intermediate
Frequency (IF) for all measurements except Direct measurements where it refers
to the Receive Frequency (RF)
I
NF: Noise Figure measured (dB)
I
Noise Temp: Noise Temperature (K) derived from measured Noise Figure
I
Gain: Gain measured (dB)