R&S
FS-K30
Remote Control
Description
of commands
Software Manual 1300.6637.42 - 03
103
DISPlay[:WINDow<1>]:TRACe<1|2>[:STATe] ON | OFF
This command switches on or off the display of the corresponding trace and
related information in the selected measurement window.
Example
DISP:TRAC ON
Switches on the display of trace 1 (Noise results).
Characteristics
*RST value: ON for both TRACe1 and TRACe2
SCPI: conforming
DISPlay[:WINDow<1>]:TRACe<1|2>:SYMBols ON | OFF
This command switches on or off the display of the symbols to mark the
measurement points for the specified trace i.e. Noise or Gain.
This command will only have a noticeable effect if the required trace is active as
set by the command DISP:WIND:TRAC<1|2>:STAT ON
.
Example
DISP:TRAC:SYMB ON
'Switches on the display of symbols for trace 1 (Noise
results)
Characteristics
*RST value: OFF for both TRACe1 and TRACe2
SCPI: device-specific
DISPlay[:WINDow<1>]:TRACe<1|2>:X[:SCALe] RF | IF
This command allows user to select RF or IF frequency to be displayed on the X
axis when the DUT is not amplifier.
The numeric suffix TRACe<1 to 2> is irrelevant.
Example
DISP:TRAC:X IF
'IF frequency will be displayed on X axis.
DISPlay[:WINDow<1>]:TRACe<1|2>:Y[:SCALe]:AUTO ON | OFF
This command switches on or off automatic scaling of the Y-axis for the specified
trace display. Automatic scaling set the Y-axis to automatically scale to best fit the
measurement results.
Note that the specified trace must be active for this command to have an
immediate effect. This can be achieved using the command
DISP:WIND:TRAC<1|2>:STAT ON.