Example
SENS:FREQ 10MHz
'the noise measurement is to be made continuously at a fixed
frequency of 10 MHz.
Characteristics
*RST value: 550 MHz
SCPI: conforming
[SENSe]:FREQuency:LIST:DATA <numeric value>,<numeric value>,<numeric
value>...
This remote control command allows a new frequency list to be specified which will
be used to perform noise and gain measurements. Each list entry will pertain to
three separate frequency entities, a RF frequency, a local oscillator frequency and
an intermediate frequency. The list arguments specified will completely overwrite
all the current frequency list entries regardless of how many entries are present
and how many entries are being supplied for the new list. The new list supplied will
remain as the active list until such time that a new list is automatically created. If
the frequency list mode is set to DIRECT then the LO and IF frequencies specified
in this command are ignored, and the query version of this command returns 0‘s
for these parameters in this case.
The values specified in this command are not used if a single frequency
measurement is initiated .
Parameter
numeric_value = a user defined set of three frequency measurements arguments
up to a maximum of 100 entry triplets in the following order. Fixed frequency, local
oscillator frequency and Intermediate frequency. The allowable frequency range is
option dependent, specified in either in Hz, kHz, MHz and GHz and will be
accurate to 2 decimal places for Hz input.
Example
SENS:FREQ:LIST:DATA 550MHz,300MHz,900MHz
'a one entry frequency list will be created with a fixed
frequency of 550 MHz, a local oscillator frequency of 300 and
an intermediate frequency of 900MHz.
Characteristics
*RST value: 550 MHz
SCPI: device-specific