1.5 Quick Start Guide
This section helps the user to quickly become familiar with R&S FS-K30 by working
step-by-step through an ordinary noise figure measurement. (Refer to section 2 for a
detailed reference guide.)
The gain and noise figure of an amplifier are to be determined in the range from 220
MHz to 320 MHz.
1.5.1 Setting up the measurement
1. Start the R&S FS-K30 application.
2. Press the
SET FREQ
softkey to open the Frequency Settings view.
3. Enter the desired frequency range in the Frequency Settings group of parameters. In
our example, enter a
Start Frequency
of 220 MHz and a
Stop Frequency
of 320 MHz.
Frequency Settings
Start Freq
220 MHz
Stop Freq
320 MHz
Step Freq
10 MHz
Mode
direct
4. Enter the desired
Step Frequency
size. In our example, 10 MHz should be entered.
Thus, a measurement is taken at 11 frequency points: 220 MHz, 230 MHz, 240 MHz,
..., 320 MHz.