List of Warnings & Error Messages
Status
reporting registers
Software Manual 1300.6637.42 - 03
135
4 List of Warnings & Error Messages
The list of possible warning & eror messages are shown bellow :
Status Bar Message
Description
Frequency list truncated, max 100 entries
The settings for start, stop and step frequencies would require a
frequency list greater than 100 entries. The list calculated is terminated
at the 100
th
entry. Try using a larger step size of splitting the test up
into a series of frequency list tests.
Missing [ENR][,|&][LossIn][,|&][LossOut] for meas.freq.
No ENR, Loss In and/or Loss Out can be determined for one or all of
the measurement frequencies. This occurs when using tables of ENR,
Loss In and/or Loss out values. Check that the frequency ranges of the
tables covers the range of frequencies to be measured. For each
measurement frequency where ENR, Loss In or Loss Out cannot be
determined 0 is used.
Generator connection error
No connection could be made to the external signal generator. Check
the connection between the analyzer and the signal generator. Also
check that the correct GPIB address has been specified for the signal
generator.
Generator not initialised
The external signal generator has not been initialised and as such
cannot be controlled by the spectrum analyzer.The signal generator
can be initialised manually by pressing the
INIT GEN
soft-key or can
automatically be initialised by selecting the
init before meas
field in the
SET MEAS
view
Generator frequency out of range
The LO frequency to be used for a measurement point is out of range
for the selected signal generator. Reduce the LO frequency or use a
signal generator with the required frequency range.