RF interference
In case of RF interferences, connect a low-pass filter to the voltage supply input of the
noise source.
The low-pass filter suppresses any interference (e.g. due to RF interference),
including interference from the supply line. This makes it possible to perform very
precise measurements.
Low pass
filter
Noise Source
0V
+28V
Noise 28V
RF
Fig. 24 Preparation for calibration
3. Set the
Second Stage Correction
parameter in the Measurement Setting view to
ON. Calibration cannot be performed until this parameter is set.
4. Start the calibration of R&S FS-K30 by pressing the
CAL
hotkey.
During calibration, the text "
Running...
" is displayed in the Status Bar at the bottom of
the screen. The progress bar indicates the progress through the calibration
measurement.
After successful calibration, the Status Bar will display “
Measurement Complete
” and
the Title Bar at the top of the screen will show a
CALIBRATED
status on the right-
hand-side.
Note that this calibration measurement calibrates only the R&S FS-K30
application and not the spectrum analyzer itself
.