R&S
FS-K30
Remote Control
Description
of commands
Software Manual 1300.6637.42 - 03
118
[SENSe]:CONFigure:MODE:SYSTem:LOSCillator FIXed | VARiable
This remote control command can be used to specify whether the local oscillator will
be used as a fixed frequency source or a variable frequency source.
The command will not have any immediate effect if a direct frequency list is being used
which is set by the remote control command SENS:CONF:MODE:DUT AMPL.
See the remote command SENS:CONF:MODE:DUT for an explanation on how the
measurement lists are created.
Parameter
FIXed
= Local Oscillator is used as a fixed frequency source
VARiable = Local Oscillator is used as a variable frequency source
Example
SENS:CONF:MODE:SYST:LOSC FIX
'the local oscillator will be set to a fixed frequency.
Characteristics
*RST value: - (SENS:CONF:MODE:DUT AMPL)
SCPI: device-specific
[SENSe]:CONFigure:MODE:SYSTem:LOSCillator:FREQuency <numeric value>
This remote control command can be used to specify a fixed local oscillator frequency
for a new frequency measurement list that requires computing. This value will form the
basis of one of the input criteria for computing a new frequency list.
The command will not have any immediate effect if a direct frequency list is being
used (which is set by the SENS:CONF:MODE:DUT AMPL command) or a variable
local osciallator frequency has been specified (which is set by the
SENS:CONF:MODE:SYST:LOSC VAR comand) .
Example
SENS:CONF:MODE:SYST:LOSC:FREQ 1MHZ
'the fixed local oscillator frequency for a new list, yet to
be created, is set to a value of 1 MHz.
Characteristics
*RST value: - 0 Hz
SCPI: device-specific
[SENSe]:CONFigure:MODE:SYSTem:IF:FREQuency <numeric value>
This remote control command can be used to specify a fixed intermediate frequency
for a new frequency measurement list that requires computing. This value will form the
basis of one of the input criteria for computing a new frequency list.
The command will not have any immediate effect if a direct frequency list is being
used (which is set by the SENS:CONF:MODE:DUT AMPL command) or a variable
intermediate frequency has been specified (which is set by the
SENS:CONF:MODE:SYST:LOSC FIX comand) .