2.5.3.4 Average
The
Average
setting is the number of measurement sweeps over which the average is
taken to produce the displayed measurement results.
The higher the number of sweeps over which the
Average
is taken, the more accurate
the measurement results will be and the more stable the display, but the measurement
time will be significantly longer.
An
Average
value of 1 means that each displayed result is produced from one
measurement sweep. This is sufficient for most cases.
2.5.3.5 RF Attenuation
RF Attenuation
is the attenuator setting that the spectrum analyzer will use for the
measurement, that is the attenuation that will be applied to the RF signal received.
To obtain a low noise figure for the analyzer and hence more accurate noise
measurements, 0 dB should be set. With high DUT power levels or for critical
matching, a higher setting is also possible. A setting of 10 dB will give a much better
VSWR of the analyzer, but will result in a worse noise figure (of the analyzer).
2.5.3.6 Automatic Ref Level
Automatic Ref Level
selects whether the reference level for measurements is
measured automatically (ON) or entered manually by the user (OFF).