R&S
FS-K30
Remote Control
Description
of commands
Software Manual 1300.6637.42 - 03
121
Characteristics
*RST value: 1
SCPI: device-specific
SOURce:EXTernal:FREQuency[:FACTor]:DENominator <numeric value>
<numeric value>
This command defines the denominator of the factor with which the analyzer
frequency is multiplied in order to obtain the transmit frequency of the selected
generator.
NOTICE!
Select the frequency offset of the generator so that the frequency range of the
generator is not exceeded with the following formula.
Numerator
F
Generator
=
F
LO
+ Offset<1>
*
Denominator
+ Offset<2>
if applied to the start and stop frequency of the analyzer.
This command is only valid in combination with option Ext. Generator Control R&S
FSP-B10.
Example
SOUR:EXT:FREQ:NUM 4
SOUR:EXT:FREQ:DEN 3
'sets a multiplication factor of 4/3, i.e. the transmit
frequency of the generator is 4/3 times the analyzer
frequency.
Characteristics
*RST value: 1
SCPI: device-specific
SOURce:EXTernal:POWer[:LEVel] <numeric value> <numeric value>
This command sets the output power of the selected generator. This command is
only valid in combination with option Ext. Generator Control R&S FSP-B10.
Example
SOUR:EXT:POW –30dBm
sets the generator level to –30 dBm
Characteristics
*RST value: 5dBm
SCPI: device-specific