R&S
FS-K30
Measurements & Settings
Measurement
Settings
Software Manual 1300.6637.42 - 03
64
Range
is the range of the DUT, that is, the maximum gain expected from the DUT.
This value is used when
Auto Ref Level
is ON and
2nd Stage Correction
is ON, to
ensure that the expected power of the measured signal will be within the optimum
operating range of the spectrum analyzer (the
Ref Level
is adjusted depending upon
the
Range
value).
To ensure accurate measurement results, the
Range
parameter should not exceed the
actual gain of the DUT by more than a margin of 10 dB.
2.5.3.9 Pre-selector
Pre-selector
is used to turn on the pre-selector when option B2 is installed.
If the B2 option is not installed, then the
Pre-selector
field is not selectable.
The
Pre-selector
provides a pre-amplifier that may be used when it has been turned on
using this parameter
Note that if both B2 and B25 options are installed only one pre-amplifier may be
switched on. In this case the setting of this parameter determines which of the Pre-
amplifiers may be switched on:
Pre-selector
ON
- Pre-amplifier for the pre-selector may be switched on.
Pre-selector
OFF
- Pre-amplifier for the electronic attenuator may be switched on.
2.5.3.10 Pre-amplifier (Pre-select)
Pre-amplifier (Pre-Select)
is used to turn on the pre-amplifier in the B2 pre-selector option.
Pre-amplifier (Pre-Select)
is only available when B2 is installed and switched on using
the
Pre-selector
setting.