R&S
FS-K30
Remote Control
Description
of commands
Software Manual 1300.6637.42 - 03
96
Example:
CALC:LIM2:LOW -30,-40,-10,-40,-30"
'Defines 5 lower limit values for limit line 2 in the preset
unit.
CALC:LIM2:LOW?
'Outputs the lower limit values of limit line 2 separated by
a comma.
Characteristics
*RST value: -
SCPI: conforming
CALCulate:LIMit<1 to 6>:LOWer:STATe ON | OFF
This command switches on or off the indicated limit line in the selected measurement
window. The limit check is activated separately with
CALC:LIM:STAT ON.
Example:
CALC:LIM4:LOW:STAT ON
'Switches on limit line 4 (lower limit)
CALC:LIM4:LOW:STAT OFF
'Switches off limit line 4 (lower limit)
Characteristics
*RST value: -
SCPI: conforming
CALCulate:LIMit<1 to 6>:LOWer:SHIFt <numeric_value>
This command shifts a limit line by the indicated value in Y-direction. The line is
shifted by modifying the individual y values
Example:
CALC:LIM3:LOW:SHIF 20DB
'Shifts all Y values of limit line 3 by 20 dB.
Characteristics
*RST value: -
SCPI: device-specific
This command is an event and is therefore not assigned an *RST value and has
no query.
3.1.3.4 CALCulate:LIMit:UPPer Subsystem
The CALCulate:LIMit:UPPer subsystem defines the upper limit line. Note that if a set
command is issued in this subsystem the limit line effected is automatically converted
to an upper limit line.