Measurements & Settings
Frequency
Settings
Software Manual 1300.6637.42 - 03
55
2.4.1.2 Stop Frequency
The
Stop Frequency
will be the frequency at which the Frequencies List stops, that is, the last
Receive Frequency (
RF
) in the list. If the number of measurement steps in the list reaches
100 steps before the
Stop Frequency
is reached, then the list generation will stop before it
reaches the
Stop Frequency
and a warning will be displayed to the user (See Section 4).
Changing the
Stop Frequency
will regenerate the list of measurement steps, using the
measurement
Mode
selected.
2.4.1.3 Step Frequency
The
Step
Frequency
will be the increment (or decrement) in Receive Frequency (
RF
) between
each measurement step in the Frequencies List. If the
Step Frequency
is larger than the
difference between the
Start
Frequency
and the
Stop
Frequency
then the Frequencies List
will contain just the
Start Frequency
and
Stop Frequency
. Changing the
Step Frequency
will regenerate the list of measurement steps, using the measurement Mode selected.
2.4.1.4 Mode
The measurement
Mode
should be selected according to the type of Device Under Test.
R&S FS-K30 provides five different measurement
Modes
:
I
Direct
I
Fixed IF, LO = RF + IF
I
Fixed IF, LO = abs(RF – IF)
I
Fixed LO, IF = RF + LO
I
Fixed LO, IF = abs(RF – LO)
Changing the
Mode
will regenerate the list of measurement steps, using the
Start, Stop
& Step Frequencies
.