R&S
FS-K30
Measurements & Settings
Measurement
results
Software Manual 1300.6637.42 - 03
52
Fig. 27 Current Value area of table of results
The left two columns display the settings being used for the measurement:
I
RF: Receive Frequency at the DUT at which the current values were measured
(Hz)
I
LO: Local Oscillator frequency (Hz) – not displayed for Direct measurements
I
IF: Intermediate Frequency (Hz) – not displayed for Direct measurements
I
ENR: ENR value (dB) – refers to the Receive Frequency (RF)
I
Loss In: Loss at the Input of the DUT (dB) – refers to the Receive Frequency (RF)
I
Loss Out: Loss at the Output of the DUT (dB) – refers to the Intermediate
Frequency (IF) for all measurements except Direct measurements where it refers
to the Receive Frequency (RF)
The right column contains the measurement results. These values will be constantly
changing allowing for monitoring of dynamic changes at this frequency.
I
NF: Noise Figure measured (dB)
I
Noise Temp: Noise Temperature (K) derived from measured Noise Figure
I
Gain: Gain measured (dB)