3.1.12.1 Equipment Settings
The following diagram shows the Equipment Settings SENSe subsystem:
[SENSe]:CORRection:ENR:MODE TABLe | SPOT
The ENR mode selection remote control command is used to specify whether a
single ENR spot value applies for all measured frequencies or whether an ENR list
table is to be used (ENR values specified at specific input frequencies).
Parameter
TABLe
= Use the ENR table
SPOT
= Use the constant ENR value
Example
SENS:CORR:ENR:MODE SPOT
'The R&S FS-K30 option uses the currently configured constant
ENR value specified for input frequencies.
Characteristics
*RST value: SPOT
SCPI: device specific
[SENSe]:CORRection:ENR:SPOT <numeric_value>
The ENR remote control command allows the ENR value applicable to the Noise
source for all input frequencies to be specified. This value provides a simple
mechanism to enter an ENR value that is applicable throughout a range of
measurement frequencies.
Example
SENS:CORR:ENR:SPOT 30
'The R&S FS-K30 option sets the internal constant ENR value
to 30 dB for all measured input frequencies. This command
will however not be effective if the remote control command
SENS:CORR:ENR:MODE:TABL was previously entered.
Characteristics
*RST value: 15 dB
SCPI: device specific