Example
SENS:CORR:LOSS:INP:TABL 1MHz,10,2MHz,12
'The R&S FS-K30 option will overwrite the current internal
loss input table with the two entry pairs specified.
Characteristics
*RST value: -
SCPI: device specific
[SENSe]:CORRection:LOSS:OUTPut:MODE SPOT | TABLe
This remote control command is used to identify whether a single output loss value
applies for all measured frequencies or whether an output loss list is to be used.
(Loss Output values specified at specific input frequencies).
Parameter
SPOT = The Constant loss output value for all measurement frequencies is used.
TABLe = The loss output table is used.
Example
SENS:CORR:LOSS:OUTP:MODE SPOT
'The R&S FS-K30 option uses the currently configured output
loss constant as the output loss value applicable for all
input frequencies to be measured.
Characteristics
*RST value: SPOT
SCPI: device specific
[SENSe]:CORRection:LOSS:OUTPut:SPOT <numeric value>
This remote control command allows the loss output constant to be set to a value
applicable for all input frequencies to be measured. This command provides a
simple mechanism to enter a loss output value that is applicable throughout a
range of measurement frequencies.
Example
SENS:CORR:LOSS:OUTP:SPOT 10
'The R&S FS-K30 option sets the internal output loss constant
value to 10 dB for all measured input frequencies. This
command will however not be effective if the remote control
command SENS:CORR:LOSS:OUTP TABL was previously entered.
Characteristics
*RST value: 0 dB
SCPI: device specific