ABL77 Service Manual
Chapter 3: Troubleshooting
Reference Table,
Continued
Error #
Message
Interpretation
Corrective Action
1306
QC Level 1 is required
before analysis can be
performed
Mandatory QC interval has
been reached. Analysis will
not be allowed until
BG/Lytes QC Level 1
analysis is performed
Perform the required
BG/Lytes QC measurement
or inactivate the required
QC interval
QC Level 2 is required
before analysis can be
performed
Mandatory QC interval has
been reached. Analysis will
not be allowed until
BG/Lytes QC Level 2
analysis is performed.
Perform the required
BG/Lytes QC measurement
or inactivate the required
QC interval.
QC Level 3 is required
before analysis can be
performed
Mandatory QC interval has
been reached. Analysis will
not be allowed until
BG/Lytes QC Level 3
analysis is performed.
Perform the required
BG/Lytes QC measurement
or inactivate the required
QC interval.
QC Level 4 is required
before analysis can be
performed
Mandatory QC interval has
been reached. Analysis will
not be allowed until
BG/Lytes QC Level 4
analysis is performed.
Perform the required
BG/Lytes QC measurement
or inactivate the required
QC interval.
QC Hct Level 1 is required
before analysis can be
performed
Mandatory QC interval has
been reached. Analysis will
not be allowed until Hct QC
Level 1 analysis is
performed.
Perform the required Hct
QC measurement or
inactivate the required QC
interval.
QC Hct Level 2 is required
before analysis can be
performed
Mandatory QC interval has
been reached. Analysis will
not be allowed until Hct QC
Level 2 analysis is
performed.
Perform the required Hct
QC measurement or
inactivate the required QC
interval.
1308
No tests remaining.
All allowable sample
analyses for the current
sensor cassette have been
performed.
Replace the sensor cassette.
1309
Hct: No End Point
Stabilization criteria not met
for Hct
Repeat measurement. If
message persists, inactivate
Hct or replace sensor
cassette.
1310
This cassette has already
been installed. Please install
a new sensor cassette.
The sensor cassette has been
identified as previously
installed.
Install a new sensor cassette.
Continued on next page